- 48 -
The device must not be placed directly beside or on top of other devices for operation.
If the medical electrical device must be placed beside or on top of other devices to oper-
ate it, it should be monitored constantly to ensure that it is operating properly in the
arrangement used.
Technical data on electromagnetic compatibility (EMC instructions) in table format are
included with the device. It can also be ordered from PARI GmbH or downloaded from
the internet at www.pari.de/en/products under "Technical Data" on the respective prod-
uct page.
6.4 Recommended safety distances
Recommended safety distances between portable and mobile high-frequency (HF) tel-
ecommunication devices and the PARI SOLE N:
The PARI SOLE N is intended for operation in an electromagnetic environment in which
radiated HF transients are monitored. The user can help to prevent electromagnetic
interference by observing the minimum safety distances between portable and mobile
HF communication equipment (transmitters) and the PARI SOLE N, as recommended
below in accordance with the maximum output power of the communication device.
Safety distance depending on transmitting frequency:
Rated power of
the transmitter
150 kHz to 80 MHz
D = (3.5/3) × SQRT(P)
80 MHz to 800 MHz
D = (3.5/3) × SQRT(P)
800 MHz to 2.5 GHz
D = (7/3) × SQRT(P)
0.01 W
0.2 m
0.2 m
0.3 m
0.1 W
0.4 m
0.4 m
0.8 m
1 W
1.2 m
1.2 m
2.3 m
10 W
3.8 m
3.8 m
7.3 m
100 W
12 m
12 m
23 m
For transmitters whose rated power does not appear in the table above, the safety dis-
tance can be calculated using the equation associated with the respective column,
where "P" is the rated power of the transmitter in Watt (W) according to the information
supplied by the device manufacturer.
6.5 Ambient conditions for operation
Ambient temperature
+10 °C to +30 °C
Relative humidity
30% to 75% (non-condensing)
Atmospheric pressure
700 hPa to 1,060 hPa