background image

Trio SCF IFU Rev E 3/31/10 2:05 PM Page 10 

Composite

C

M

Y

CM

MY

CY CMY

K

QUESTIONS ABOUT TRIO? 1-888-335-6946

   

|

  

9

H

TRIO NEBULIZER HANDSET CLEANING (CONT)

3. To wash the Trio Nebulizer Handset, start by thoroughly rinsing

each of the disassembled parts under warm running tap water for
approximately 10 seconds each.

Rinse the Aerosol Head for at least 10 full seconds on each side (Figure
26). Never use a brush or any other object to clean the AEROSOL HEAD
(Figure 27). THOROUGH RINSING OF BOTH SIDES OF THE AEROSOL
HEAD HELPS TO PREVENT CLOGGING AND ENSURE PROPER OPERATION.

4. Add a few drops of clear liquid dish soap to a clean bowl or basin and

add warm tap water (Figure 28).

5. Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under

running, warm tap water (Figure 29).

If Disinfecting after Cleaning, skip to section I and follow the steps.

6. Air dry the Trio Nebulizer Handset parts on a clean, lint-free towel

(Figure 30).

7. After the Trio Nebulizer Handset parts are completely dry place them

in a bag for storage. Re-assemble the Trio Nebulizer Handset just prior
to the next treatment.

8. If the Controller needs cleaning, wipe down with a slightly damp, soft

cloth (Figure 31). DO NOT allow liquid to get inside the Controller.

26

Figure

27

Figure

28

Figure

30

Figure

31

Figure

                       CONTROLLER

29

Figure

Summary of Contents for Trio

Page 1: ...ral law restricts this devicetosalebyoronorderofaphysician Read and understand these Instructions for Use and all safety precautions it contains Improperusecancauseserious or fatal injury and or illne...

Page 2: ...on and safety precautions for the Trio Electronic Nebulizer Save these Instructions For Use for future reference BeforeusingyourTrioElectronicNebulizer youshouldreadandunderstand the entire Instructio...

Page 3: ...NEVER operate the Controller if it is improperly or incompletely assembled or damaged Section G Taking a treatment shows the alerts to these situations when the Trio Electronic Nebulizer is improperl...

Page 4: ...o not have all the items please call the number listed below FOUR AA BATTERIES MEDICATION RESERVOIR AEROSOL CHAMBER MOUTHPIECE WITH EXHALATION VALVE AC POWER SUPPLY AEROSOL HEAD INHALATION VALVE CARRY...

Page 5: ...eries will provide approximately 2 hours of continuous treatment You may also use AA rechargeablebatteries Pleasesee Specifications ontheback cover of these Instructions For Use NOTE IT IS STRONGLY RE...

Page 6: ...fingers Figure 8 F TRIO NEBULIZER HANDSET ASSEMBLY WARNING TheTrioNebulizerHandsetisnotsterile CleanandDisinfecttheTrio NebulizerHandsetbeforethefirstuse seeSectionHforCleaningand SectionIforDisinfect...

Page 7: ...withTrioElectronic Nebulizer a Trio Nebulizer Handset Attached To Controller Place the Controller on a clean flat stable surface Slide the Trio Nebulizer Handset into the Controller cradle as far as i...

Page 8: ...en delivered a You will hear the Treatment Complete Alarm 2 beeps and 2 green lights b The Trio Electronic Nebulizer will automatically shut off 10 If for any reason you must stop the treatment before...

Page 9: ...bulizerHandsetimmediatelyaftereveryuse Donot wash the Controller Nebulizer Connection Cord AC Power Supply or PlugConverters Useliquidsoapdesignedforwashingdishes Donot use white liquiddishsoaps e g I...

Page 10: ...f clear liquid dish soap to a clean bowl or basin and add warm tap water Figure 28 5 Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under running warmtapwater Figure 29 If Disinfect...

Page 11: ...ain after the treatmentintheMedicationReservoir Ifmorethanafewdropsremain cleantheAerosolHead SectionH If afterpropercleaning morethan afewdropsremainintheMedicationReservoir call1 888 335 6946 QUESTI...

Page 12: ...batteries following the procedure in Section D Please note that the Controller has a built in automatic shutoff after 10minutes anditwillturnoffatthattime Ifthissituationcontinues three or four times...

Reviews: