FREQUENTLY ASKED QUESTIONS AND

TROUBLESHOOTING (CONT)

QUESTION 4:

How long will a new set of batteries last?

Answer:

  A new set of batteries will provide approximately 2 hours of

continuous treatment.

Answer:

  Consider using rechargeable batteries.

QUESTION 5:

What if no mist is coming out of the Trio Nebulizer Handset?

Answer:

 First, check that the Controller has power. Secondly, make

sure the Nebulizer Connection Cord is correctly attached. Thirdly,
check to be sure that the Trio Nebulizer Handset has been assembled
properly (Section F). Lastly, check to be sure medication is in the
Medication Reservoir. If the Trio Electronic Nebulizer still does not
function properly, call 1-888-335-6946.

QUESTION 6:

What if liquid is leaking from the Trio Nebulizer Handset during

the treatment?

Answer:

 Check to see if the Inhalation Valve has been warped

or deformed.

Answer:

 Be sure the Inhalation Valve is properly attached to the Aerosol

Chamber to prevent leaks (Section F, 6a and 6b).

Answer:

 Ensure the Aerosol Head is inserted correctly (Section F, 7,

7a, 7b). Next, ensure the Medication Reservoir and the Aerosol
Chamber are closed correctly (Section F, 9).

QUESTION 7:

  What if my Controller shuts off before my treatment begins,

or does not restart?

Answer:

 Press the On/Off button for one full second. If it still shuts off

after several (3-4) attempts, call 1-888-335-6946.

QUESTION 8:

  What if my Controller does not shut off at the end of

my treatment?

Answer: 

Press the On/Off button again.

Answer: 

Disconnect the AC Power Supply from the Controller and

remove the batteries. Then, reinsert the batteries following the
procedure in Section D.

Please note that the Controller has a built-in, automatic shutoff after

10 minutes, and it will turn off at that time. If this situation continues
three or four times, call 1-888-335-6946.

L

SPECIFICATIONS

Aerosol Head Replacement............................................ Every 90 days

Weight: Controller and Trio Nebulizer Handset ............................. 6 oz

Weight: Controller, Trio Nebulizer Handset, and Batteries ....... 9.5 oz

Dimensions of Trio Nebulizer Handset .................. 1.93" X 2.5" X 5.5"

Electrical Requirements ........................... 110 V - 240 V, 50 Hz/60 Hz

Power Wattage ......................................... 2.0 Watts under normal load

TRANSPORT / STORAGE

Ambient Temperature ........................................................ -4

°

 to 140

°

 F

Ambient Humidity .................................................................. 10% to 95%

RECHARGEABLE BATTERIES

Rechargeable Batteries ..................................... 1.2V-1.5V, min 1500mAh

Rechargeable Battery Operation .............................. About 2 hours of

continuous treatment

Rechargeable Battery Charge Time ................................ up to 15 hours

(depends on type of battery charger)

DISPOSABLE BATTERIES

Panasonic Power Max 3 or equivalent ............................................ 1.5V

Panasonic Power Digital or equivalent ............................................ 1.5V

MANUFACTURED IN GERMANY & DISTRIBUTED BY:

PARI Respiratory Equipment, Inc.

SourceCF

®

, a subsidiary of

2943 Oak Lake Blvd.

Eurand Pharmaceuticals, Inc.

Midlothian, Virginia USA

Huntsville, Alabama 35806 USA

1-800-FAST-NEB (327-8632)

1-888-335-6946

PATENT NO.S:

US 5,152,456; US 5,261,601; US 5,518,179; US 6,983,747;
US 6,962,151 Additional Patents Pending
Canadian Patent CA 2,389,936

DESIGN AWARDS:

©2010 PARI Pharma GmbH. All rights reserved. The PARI Pharma logo is a registered trademark, and
Trio is a registered trademark of PARI Pharma GmbH. SourceCF® is a registered trademark of SourceCF,
a subsidiary of Eurand Pharmaceuticals, Inc. Touchspray® Technology is made under license from
the Technology Partnership, plc. Control III is a registered trademark of Maril Products Inc.

078D1004 Rev E 03/10

Summary of Contents for Trio

Page 1: ...ral law restricts this devicetosalebyoronorderofaphysician Read and understand these Instructions for Use and all safety precautions it contains Improperusecancauseserious or fatal injury and or illne...

Page 2: ...on and safety precautions for the Trio Electronic Nebulizer Save these Instructions For Use for future reference BeforeusingyourTrioElectronicNebulizer youshouldreadandunderstand the entire Instructio...

Page 3: ...NEVER operate the Controller if it is improperly or incompletely assembled or damaged Section G Taking a treatment shows the alerts to these situations when the Trio Electronic Nebulizer is improperl...

Page 4: ...o not have all the items please call the number listed below FOUR AA BATTERIES MEDICATION RESERVOIR AEROSOL CHAMBER MOUTHPIECE WITH EXHALATION VALVE AC POWER SUPPLY AEROSOL HEAD INHALATION VALVE CARRY...

Page 5: ...eries will provide approximately 2 hours of continuous treatment You may also use AA rechargeablebatteries Pleasesee Specifications ontheback cover of these Instructions For Use NOTE IT IS STRONGLY RE...

Page 6: ...fingers Figure 8 F TRIO NEBULIZER HANDSET ASSEMBLY WARNING TheTrioNebulizerHandsetisnotsterile CleanandDisinfecttheTrio NebulizerHandsetbeforethefirstuse seeSectionHforCleaningand SectionIforDisinfect...

Page 7: ...withTrioElectronic Nebulizer a Trio Nebulizer Handset Attached To Controller Place the Controller on a clean flat stable surface Slide the Trio Nebulizer Handset into the Controller cradle as far as i...

Page 8: ...en delivered a You will hear the Treatment Complete Alarm 2 beeps and 2 green lights b The Trio Electronic Nebulizer will automatically shut off 10 If for any reason you must stop the treatment before...

Page 9: ...bulizerHandsetimmediatelyaftereveryuse Donot wash the Controller Nebulizer Connection Cord AC Power Supply or PlugConverters Useliquidsoapdesignedforwashingdishes Donot use white liquiddishsoaps e g I...

Page 10: ...f clear liquid dish soap to a clean bowl or basin and add warm tap water Figure 28 5 Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under running warmtapwater Figure 29 If Disinfect...

Page 11: ...ain after the treatmentintheMedicationReservoir Ifmorethanafewdropsremain cleantheAerosolHead SectionH If afterpropercleaning morethan afewdropsremainintheMedicationReservoir call1 888 335 6946 QUESTI...

Page 12: ...batteries following the procedure in Section D Please note that the Controller has a built in automatic shutoff after 10minutes anditwillturnoffatthattime Ifthissituationcontinues three or four times...