ELECTROMAGNETIC COMPATIBILITY – Guidance and
manufacturer's declaration DIN EN 60601-1-2:2007
(IEC 60601-1-2:2007)
Electromagnetic compatibility
Medical electrical devices are subject to special precautionary measures in particular
regarding EMC with the installation and operation.
Portable and mobile HF-communication devices e.g. mobile phone can affect medical
electrical devices.
Use of accessories and lines other than those indicated can lead to increased
transmission or reduced noise immunity of the equipment. The equipment has to be
operated exclusively with original accessories.
The device should not be used beside or on top of other equipment. If such use is
unavoidable, the device should be observed to verify normal operation in the
configuration in which it will be used.
The EUT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EUT must ensure that it is used in such environment.
There is no essential performance.
Electromagnetic emissions
Emission test
Compliance
Electromagnetic environment –
Guidance
RF emissions CISPR 11
Group 1
EUT uses RF energy only for its internal
function.
Therefore, RF-emission is very low and not
likely to cause any interference with nearby
electronic equipment.
Class B
The EUT is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Passed
Table 1 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)