background image

Shut down 

 

CAUTION: ENSURE HEAD WILL NOT CONTACT MICROSCOPE!! 

 

1) 

DO NOT

 

hit 

“Lift Z” 

right away.

  

The stage will hit the microscope on the way up and 

damage one or both of them. Use Z-stage control to lift tip away from sample a small distance so 
the tip is no longer touching the sample.   

2) Open the acoustic enclosure 

3) Enable Focus Follow and then hit ‘Lift Z’ button in Z-stage control. Both the microscope and 
Z-stage will raise. 

4) Refocus microscope on cantilever as needed.

 

5) Turn off laser switch, unplug head, and remove head. 

5a) If the tip you used was good, leave it on the AFM. Otherwise, remove head and place tip in 
the “Bad Probes” box 

6) Remove sample from stage and replace head, plug head in. 

7) Close XEP/XEI/XEC software 

8) Turn off Light Bank  

9) Leave the SPM Controller on 

 

**Log tip usage, clean up after yourself, and disable in Badger** 

 

 

 

 

 

 

 

 

 

 

Note: We recycle used sample pucks and also used tips.  Please put used items in the proper 
places.  Do not throw them in the trash. 

 

If you notice any problems while running the instrument, please make a comment on Badger and 
email the lab managers at nano-spmstaff@lists.stanford.edu. 

Summary of Contents for XE-100

Page 1: ...ser will move on the quadrant photodetector This is relayed to a motor controlling the tip which adjusts the tip s vertical position until the laser is aimed in the center of the quadrant photodetecto...

Page 2: ...ting tip and place it in the good tips box Ensure new tip is rested in place and that bearings fit into tip slots 4 Replace head and push flaps away from you until slightly tight This secures the head...

Page 3: ...n Look at where the light from the microscope is shining on the tip assembly as this will help you to center the microscope Position laser on cantilever If you don t see the laser reflected off the ca...

Page 4: ...e appropriate value If after several tries you still can t get the sum high enough remove the head and check mirror 2 s position visually it should be roughly parallel to the base Adjust mirror 2 knob...

Page 5: ...o 1 00 Cantilever General 4 Click OK 5 Turn Head back on in software Frequency Sweep Window will open 6 The software will automatically detect the cantilever s resonance peak and select a driving freq...

Page 6: ...B and C D are still less than 0 3 V 9 Shut acoustic enclosure 10 Click Approach button in Z stage control When the approach stops the light in the Z stage Control window will stop blinking Ensure tha...

Page 7: ...will change the distance the tip hovers above the sample surface when scanning and can improve resolution of features surface tracking This can be done manually by changing the number in the Set Poin...

Page 8: ...h XEI You can send images directly to the XEI data analysis software from the Buffer in XEP by right clicking on them Or you can open them from the spmdata or User Files folders Flattening your sample...

Page 9: ...er as needed 5 Turn off laser switch unplug head and remove head 5a If the tip you used was good leave it on the AFM Otherwise remove head and place tip in the Bad Probes box 6 Remove sample from stag...

Page 10: ...ion on center of photodetector A B 3 5V A B and C D 0 3V 6 Select appropriate mode scanner and frequency 7 Approach surface i Get optically close several mm ii Focus microscope 50 100um from surface m...

Reviews: