background image

70-PHW-1025 R0.6 023-037X MT-IR Refrigerant Leak Detector spec & Install.doc 

10 

CALIBRATION 

Prior to shipment, all sensors manufactured are factory calibrated. The calibration method will set 
the base level (or Zero Level) and the gain (or Slope). As the sensor gets older or the ambient 
condition changes drastically, the Zero Level may drift upward or downward. The Gain (Slope) will 
not normally change. 
An adjustment of the Zero Level is necessary to ensure that the MT-IR is reading accurately. This 
should be done every 6 months. This function can be performed by conducting the Calibration Push 
button. 
PUSH BUTTON 1 -- Located just to the left of the red DIP Switch block. 
1) Make sure the unit is properly wired up and all dip switches are in the off position. 
2) Allow at least three hours warm up time. The ambient temperature should be stable and within 
the desired operating range. 
3) Make sure no refrigerant gas is present during calibration. 
4) Depress and hold down the push button 1 (located to the left of DIP SW1) for at least 8 seconds 
until the calibration LED red light above the push button lights. Release the push button. 
5) The LED red light will go off. If not, make sure switch 6 is off. Consult factory if calibration 
LED light stays on or keeps flashing after calibration or during normal operation. 
NOTE: If the sensor is reading greater than 80, the sensor will not calibrate. Check for gas leak 
around the sensor. 
 

MEASURED GAS CALIBRATION 

Equipment Required. 
1 Refrigerant Sensor Calibration and Test Kit which includes: 
a. Flow Regulating Valve - 1 Liter per minute 
b. Calibration Gas Canister - 79.1% Nitrogen / 20.9% Oxygen (Zero Gas) 
c. Calibration Gas Canister - Nitrogen Balance / Measured quantity of Refrigerant Gas (Span Gas) 
d. Vinyl Tubing for IR Sensor. 
The measured gas calibration procedure may be used as a double check of sensor accuracy and only 
needs to be done if there are questions of whether or not the sensor is working. It is not necessary 
to use measured gas to calibrate the infrared sensor. This procedure involves changing the air 
currents drastically inside the sensing chamber, so please use the procedure listed below only as a 
test. Since the gas is forced into the sensor the reading will not be as accurate as a normal reading. At 
the factory we allow for normal airflow that would occur in a real leak when we calibrate the sensor 
and set the zero level. So there will be an artificial reading between the zero gas and measured gas 
because of the unusual air currents involved. 
1) Place the Plastic tubing that comes with the calibration kit over the inlet valve located to the right 
of the wiring plug and connect the other end to the calibration gas canister regulator. 
2) Run the 20.9% O2 gas into the sensor for about 5 minutes and wait for the reading to settle. 
Press the push button on the sensor right above the dipswitch for 10 seconds. This will be the 
artificial zero level. 
3) Next run the measured refrigerant gas into the sensor for about 5 minutes. The difference 
between the artificial measured gas reading and the artificial zero level reading should be close to the 

Summary of Contents for 023-037X

Page 1: ...in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without the prior written permission of Micro Thermo Technologies 1997 2015 b...

Page 2: ...librated to monitor for a specific refrigerant gas There are two models of IR sensor for each gas depending upon the monitoring environments Machine room for most applications above freezing and Cold...

Page 3: ...r refrigerants mount 18 to 24 from the floor For Ammonia mount on or near the ceiling CALIBRATION Zeroadjustmentrecommendedatsixmonth intervals OPTIONALRELAY RemoteSPDTOutput1setto500ppm RemoteSPDTOut...

Page 4: ...70 PHW 1025 R0 6 023 037X MT IR Refrigerant Leak Detector spec Install doc 3 MOUNTING INSTRUCTIONS...

Page 5: ...t the sensor can be determined by lighting a match close to the sensor If the match is blown out by the air current mounting the sensor inside a pull box with the knockouts opened slightly or some oth...

Page 6: ...feet with the locally available wall mounting hardware Pole Mounting The sensor can be mounted to a support pole by utilizing Uni Strut type C braces Attach the C braces to the back of the housing per...

Page 7: ...sequencer input terminal block Do not connect the wire of the 24VDC to the relay sequencer This will prevent ground loop to affect signal accuracy Default relay threshold settings are 500ppm and 1000...

Page 8: ...23 0280 1000ppm model as they have compatible settings 3 Open the newly dropped icon and then select the Refrigeration Leak alarm settings model in the Pick Alarm Settings window of the Alarm Settings...

Page 9: ...o long term zero compensation It will recalibrate up to 10 ppm every 7 days No more than 100 ppm between manual calibrations Disable for low level detection On zero Compensation SW4 OFF Disabled SW4 O...

Page 10: ...microprocessor error LED L2 is used for zero calibration When pushbutton 2 is pressed LED 2 will light after 8 seconds and turn off once the button is released Status Condition Off Normal operation ON...

Page 11: ...ment Required 1 Refrigerant Sensor Calibration and Test Kit which includes a Flow Regulating Valve 1 Liter per minute b Calibration Gas Canister 79 1 Nitrogen 20 9 Oxygen Zero Gas c Calibration Gas Ca...

Page 12: ...fter power up 5 Variation between 1 to 10 ppm is Normal It is not gas detection this is to shows the sensor is powered 6 Calibration and Zeroing of Sensor The Infrared Sensor is factory calibrated how...

Page 13: ...L1 on Flashing Switch 5 is on to force a 200 ppm reading on the sensor Microprocessor detects error in the sensor signal Push Dip Switch 5 down If switching SW5 does not work call MTT technical suppor...

Reviews: