Press button 4, inp 2 appears, press button 4, and it should read 20.0
12)
Press button 4, dspl 2 appears, press button 4, using buttons 2, & 3 will change the number
and position change display until it reads 68.0.
13)
Press button 4 to finish and save, if the probe is attached the display will change to a reading
and after approximately 10 seconds flash to real dew point. If the probe is not attached it
should flash -199.9.
If display continues to read -199.9 the following are possibilities:
Input selected in setup menu
TC/RTD switch position. (Dip switch on back of module)
Corresponding signal at signal connector
6.
Probe Recalibration
Routine maintenance of the DB1000 probe / transmitter is confined to regular
re-calibrations. This work can only be done by exposure of the moisture
sensor to sample gases of known moisture content. The manufacturer of the
probe provides calibration services traceable to NIST. Contact Airtek for
further details.
7.
Warranty
The DB1000 is equipped with a 1-year parts and labor warranty. The DB1000
is guaranteed to be free from defects in material and workmanship for a period
of 1-year from date of shipment. This warranty does not include recalibration
of probe or meter. For the warranty to be valid, the customer must call our
service dept. and speak with a service representative. No warranty claims will
be honored without first speaking with an Airtek customer service
representative.