background image

10

Parker Hannifin Corporation

 Bulletin TI-HFX-3D

Maintenance

To avoid system damage and/or personal harm, isolate the Balston Nitrogen Generator from

To avoid system damage and/or personal harm, isolate the Balston Nitrogen Generator from

To avoid system damage and/or personal harm, isolate the Balston Nitrogen Generator from

To avoid system damage and/or personal harm, isolate the Balston Nitrogen Generator from

To avoid system damage and/or personal harm, isolate the Balston Nitrogen Generator from
the compressed air supply and fully depressurize prior to performing any maintenance

the compressed air supply and fully depressurize prior to performing any maintenance

the compressed air supply and fully depressurize prior to performing any maintenance

the compressed air supply and fully depressurize prior to performing any maintenance

the compressed air supply and fully depressurize prior to performing any maintenance
activities.

activities.

activities.

activities.

activities.

All maintenance activities for the Balston HFX-3 and HFX0-3 Nitrogen Generators should be

All maintenance activities for the Balston HFX-3 and HFX0-3 Nitrogen Generators should be

All maintenance activities for the Balston HFX-3 and HFX0-3 Nitrogen Generators should be

All maintenance activities for the Balston HFX-3 and HFX0-3 Nitrogen Generators should be

All maintenance activities for the Balston HFX-3 and HFX0-3 Nitrogen Generators should be
performed by suitable personnel using reasonable care.

performed by suitable personnel using reasonable care.

performed by suitable personnel using reasonable care.

performed by suitable personnel using reasonable care.

performed by suitable personnel using reasonable care.

Required maintenance for the Balston Nitrogen Generation System consists of changing the prefilter
and final filter cartridges (annually), changing the galvanic cell in the oxygen analyzer (annually),
checking the performance of the system, and calibrating the oxygen analyzer.  The recommended
service schedule and replacement part numbers are outlined at the end of this section.  The calibra-
tion procedure for the oxygen analyzer is outlined on page 6.  The analyzer should be calibrated on
a regular basis, as dictated by the application.

Changing filter cartridges more frequently will translate into direct energy savings and reduced
operating costs.  Annual electricity costs to operate a typical 100 HP compressor can be as high as
$50,000.  Pressure drop in the system adds to this expense.  A system operating at 100 psig that is
exeriencing a 2 psig pressure drop through a filter, requires an additional 1% in operating energy
costs or approximately $500.00+ per year.

Replacement prefilter cartridges, final membrane filter cartridges, and galvanic cells may be ordered
through your local representative.  For convenience, these replacement items have been packaged
into a maintenance kit for each generator.  Maintenance Kits provide a 6 month supply of filtration
replacement components.

The only tools required to change the prefilter cartridges and the final membrane filter cartridge is a
Phillips head screwdriver.  All filter housings which require routine service are easily accessible
when the front filtration access panel is opened.  (Remove the four corner screws.)  The two coalesc-
ing prefilters are Balston 2002N-1B1-DX and 2002N-1B1-BX filter assemblies (see Figure 7).  Re-
place these filter cartridges with Balston 100-12-DX and 100-12-BX filter cartridges.  When replacing
the prefilter cartridges, be sure to install the proper grade filter in the proper housing (the housings
are labeled).  The final membrane filter is a Balston 2002N-0A0-95 filter assembly.  Replace this
membrane filter cartridge with a Balston Grade GS 100-12-95 membrane filter cartridge.

The filter cartridges in the filter assemblies are removed by:  twisting the filter bowl clockwise 1/8 of a
turn while pushing up gently, lowering the filter bowl away from the filter head, and unscrewing the
element retainer from the base of the cartridge.  Insert the new filter cartridge and reassemble the
housing in reverse order.

The time required to service all three stages of filtration is less than 15 minutes.

Filter Cartridge
Replacement

!

Summary of Contents for Balston HFX-3

Page 1: ...separate air into a concentrated nitrogen output stream and an oxygen enriched permeate stream To monitor the purity of the nitrogen stream Parker has incorporated an oxygen analyzer into the design o...

Page 2: ...ure 1 Air separation takes place in the membrane module This module consists of bundles of hollow fiber membranes The compressed air enters the center bore of these fibers and travels the length of th...

Page 3: ...junction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times systems to assure proper nitrogen purity levels at all times systems to assure proper nitrogen...

Page 4: ...nt surrounding the nitrogen generator should also be adequately ventilated The environment surrounding the nitrogen generator should also be adequately ventilated The generator creates a 30 to 40 oxyg...

Page 5: ...sure to maintain the proper polarity black red The HFX0 3 Nitrogen Generator is preset at the factory for operation at 120 VAC The voltage setting for the generator is shown through a small window on...

Page 6: ...outside circuitry alarm relay contacts to outside circuitry alarm relay contacts to outside circuitry alarm relay contacts to outside circuitry The relay contacts are rated for 250 VAC 5 amps resistiv...

Page 7: ...gen using a span gas known quantity of oxygen per gas supplier or compressed air 20 9 oxygen In the single point method only one point in the calibration range is set using either span gas or compress...

Page 8: ...oxygen analyzer on a routine basis If the nitrogen purity level falls below the desired level readjust the flow control valve until the proper nitrogen purity level is reached Note If more than minor...

Page 9: ...vered from the system If the temperature difference is 10 F 20 F 5 C 11 C this equilibrium period may be as long as 60 minutes The inlet air tempera ture and more importantly inlet air dewpoint must n...

Page 10: ...vice schedule and replacement part numbers are outlined at the end of this section The calibra tion procedure for the oxygen analyzer is outlined on page 6 The analyzer should be calibrated on a regul...

Page 11: ...en servicing This equipment has fuses in both neutral and phase lines Use care when servicing Occasionally one or both of the fuses P N 13221 in the generator may burn out The fuses are located in the...

Page 12: ...ve 95 F 35 C the longevity of the membrane will be reduced and the warranty will be void 6 6 6 6 6 The drain lines from the first two stages of filtration should be piped away to an appropriate collec...

Page 13: ...58 F 50 C Particles 0 01 m None None Suspended Liquids None None Commercially Sterile Yes Yes Max Operating Pressure 145 psig 10 barg 145 psig 10 barg Maximum Pressure Drop 10 psi 0 7 bar 10 psi 0 7 b...

Page 14: ...d range during calibration Replace sensor P N 72695 Operating Conditions To arrange for system service contact the Technical Services Department at 1 800 343 4048 8AM to 5PM Eastern Time For location...

Page 15: ...cription Description Description Description Caution refer to accompanying documents for explanation Refer the caution warning note indicated for explantion Caution risk of electric shock 3 3 3 3 3 Ex...

Page 16: ...rhill MA 01835 0723 Tel 978 858 0505 Fax 978 556 7501 Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation F...

Reviews: