2.1 Technical Specification
This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.
Units
LCMS15-0
LCMS15-1
LCMS20-0
LCMS20-1
LCMS30-0
LCMS30-1
LCMS40-0
LCMS50-0
Drying, Sheath and Nebulisation Gas
Flow
L/min
15
20
30
40
50
Outlet Pressure
bar g
(psi g)
7
(101.5)
Purity
(with respect to O
2
)
%
>99.5
>99
>98
>99
>98
Inlet Parameters
Integral Compressor
-
N/A
Yes
N/A
Yes
N/A
Yes
N/A
Inlet Air Quality
ISO 8573-1: 2010
1.-.1
N/A
1.-.1
N/A
1.-.1
N/A
1.-.1
Inlet Air Pressure
bar g
(psi g)
8.5
(123.3)
N/A
8.5
(123.3)
N/A
8.5
(123.3)
N/A
8.5
(123.3)
Maximum Inlet Air
Temperature
o
C
(
o
F)
35
(95)
N/A
35
(95)
N/A
35
(95)
N/A
35
(95)
Flow
(@ 8.5 bar g)
L/min
70
N/A
70
N/A
130
N/A
130
Port Connections
Air Inlet
1
1.
Compression fitting
-
1/4
N/A
1/4
N/A
1/4
N/A
1/4
Gas Outlet
1
-
1/4
2
2.
Using the 1/8 - 1/4 adaptor provided.
Electrical Data
Supply Voltage
Vac
230 ± 10% 50Hz
Connection type
-
IEC320
Current
3
3.
The power consumption when in standby mode is 35W for all models.
A
0.8
5.8
0.8
5.8
0.8
5.8
0.8
Fuse
A
2.0
3
6.3
4
2.0
3
6.3
4
2.0
3
6.3
4
2.0
3
Supply Voltage
Vac
120 ± 10% 60Hz
Connection type
-
IEC320
Current
3
A
1.6
11.6
1.6
11.6
1.6
11.6
1.6
Fuse
A
2.0
4
4.
(Anti surge (T), 250V, 5 x 20mm HBC, Breaking Capacity 1500A @250V, IEC 60127, UL R/C)
12.5
5
5.
(Anti surge (T), 250V, 6.3 x 32mm HBC, Breaking Capacity 1500A @440V, IEC 60127, UL R/C)
2.0
3
12.5
4
2.0
3
12.0
4
2.0
3
Environmental Data
Ambient Temperature
o
C
(
o
F)
5–40
(41–104)
Relative Humidity
50% @ 40
o
C (80%
MAX
< 31
o
C)
IP Rating
IP20, NEMA 1, indoor use only
Pollution Degree
2
Installation Over voltage
Category
II
Maximum Altitude
m
(ft.)
2000
(6562)
Noise
dB(A)
< 65