IntroDUCtIon
Parker Hannifin icountACM202024
g
4
Introduction
The icountACM202024 Aviation Contamination Monitor
from Parker represents the most up-to-date technology in
automatic particle counting, and is the first truly portable
monitor.
The icountACM202024 has been developed for
monitoring contamination in Avtur and other Hydrocarbon
fuels, in accordance with the Energy Institute (EI) method
IP 564.
The icountACM202024 is a complex instrument, but
at the same time has reliability, simplicity and ease of
operation designed-in.
This User Guide can show you step-by-step how to get
started, how to obtain reliable measurements, and how to
interpret the results.
The real benefits to be gained from the icountACM202024
are achieved through regular use, particularly as an
effective comparator.
With a typical test taking only two minutes, the
icountACM202024 is ideal for use as your standard fluid
contamination monitoring instrument.
Above all, the icountACM202024 has been designed for
practical use in real-world environments.
Features
Operational features
Test time
2 minutes, with a repeat test time of every 2 minutes
Principle of operation
Optical scanning analysis and measurement of actual particulates and inference to
water presence (as per Clear and Bright methods)
Particle size ranges
4+, 6+, 14+, 21+, 25+ and 30+ microns
ISO code range
7–22
Data entry
32 character two line dot matrix LCD. Full alphanumeric entry facility on keypad.
Data retrieval
Memory access gives test search facility
Communications
Interface via RS-232 connection at 9600 baud rate
Printer
Integral 16-column printer for hard copy output
Calibration
Each unit is individually tested and calibrated by online methods, in accordance with
ISO procedures.
Max. working pressure
420 bar
Max. flow rate
380 l/min when used with System 20 Sensors.
Memory capacity
300 test (scrolling memory) capacity
Computer compatibility
Interface via RS232 connection @ 9600 baud rate
Portability
Weighs 6.5kg, including the built-in battery
Power requirement
12Vdc supplied by rechargeable or replaceable battery module
System connection
Via System 20 Inline Sensors
Leak-free sampling
System 20 Sensors ensure sealed fluid extraction and no contamination ingress
Certification
This product complies with all relevant EC declarations of conformity.
Fail-safe features
Special diagnostics are incorporated into the icountACM202024 microprocessor control to ensure effective testing.
Circuitry
Incorporates an internal diagnostic programme to ensure integrity of results
Adequate flow
Flow test facility ensures adequate flow
Data management
The specially designed
ParSmart Downloader
software package is supplied to facilitate the downloading of test
results to a computer.
Summary of Contents for icountACM202024