background image

Parker Hannifin Manufacturing (UK) Ltd

24

USA

Getting started

icountBSplus setup

The icountBSplus is supplied with an integral touch-screen and stylus.

CAUTION: Do not use any sharp items on the touch screen as this may damage the 

product and is not covered by any warranty. If in doubt please refer to the operator 

manual or contact Parker Hannifin Manufacturing (UK) Ltd for advice.

See ‘Electrical connections’  

pages 18–19

See ‘Fluid connections’  

(if applicable) pages 20–21

1

2

Turn on unit

Panel button illuminates 

(flashes when charging)

3

4

Locate stylus

Remove stylus

Summary of Contents for icountBSplus

Page 1: ...icountBSplus icountBSplus Bottle Sampler Manual DD0000014_US RevA 2013 Parker Hannifin Manufacturing UK Ltd www parker com hfde USA...

Page 2: ...s 15 Battery charging status icons 16 Contamination Standards 17 Setting up 18 Using the Touch Screen 18 Electrical connections 18 Fluid connections 20 Printer paper installation 22 Getting started 24...

Page 3: ...ruption With its Innovative touch screen technology integrated battery power source and secure safe operating parameters the new icountBSplus will deliver the performance characteristics by meeting ac...

Page 4: ...concerns SERVICE REPAIR Any servicing or repair work must be carried out by a Parker approved service center Contact your local Parker Hannifin sales office for recalibration services WARNING A red ba...

Page 5: ...ct Parker Hannifin for advice CAUTION Disconnect from Mains supply before accessing the battery compartment Do not dismantle short circuit or modify batteries in any way Do not charge the battery outs...

Page 6: ...not open the unit To ensure compliance with the certification users are NOT permitted to open the unit under any circumstances Doing so will invalidate the unit s calibration and it would NOT subseque...

Page 7: ...icountBSplus Bottle Sampler Manual 7 USA EC declaration of conformity...

Page 8: ...Y 1 Part number reference 2 Serial number 3 Date of manufacture D O M 4 Country of manufacture 5 Refer to handbook 6 Waste electrical and electronic equipment WEEE directive End of life disposal is th...

Page 9: ...t the volume of the test sample and the number of measurements taken for each bottle sample The results are reported immediately but also stored in the unit s internal memory and may be downloaded lat...

Page 10: ...0ml Selectable Flush Sample Sizes 10 15 20 25 50 and 100ml Selectable number of samples taken in one time dependant on the size of the volume and flush sizes selected 1 2 3 4 or 5 tests A low cost sol...

Page 11: ...supply 3 bar maximum 6 Battery compartment 7 USB connections A and B 8 Mains On Off switch must be switched on to charge battery 9 Ventilation fan DO NOT BLOCK Mounting Best practice mounting orientat...

Page 12: ...Parker Hannifin Manufacturing UK Ltd 12 USA Dimensions Dimensions are given in mm inches Pneumatic hydraulic circuit 393 3 15 5 405 7 16 194 7 6 207 5 8 2 525 2 20 7...

Page 13: ...amination Standards DD0000015 on CD ACFTD ISO 4406 1987 ISO 4406 1991 NAS 1638 AS4509E Diff AS4509E Cum GOST 17216 2001 MTD ISO 4406 1999 NAS 1638 AS4509E Diff AS4509E Cum Channel sizes Channel Sizes...

Page 14: ...to 70 C 4 F to 158 F Sample bottle size Parker Standard Bottle 250ml for use in icountBSplus ACC6NW001 for 2 bottle pack ACC6NW002 for 50 bottle pack see page 46 48 Flush sample size Selectable option...

Page 15: ...Screen calibration Registration Language selection System Information User Configuration Help Update software Parker only Calibration check Parker only 7 Default setting reset see page 14 8 Return to...

Page 16: ...rcuit fault 100 charge 100 Tests Shop Air 64 Tests Battery 80 charge 86 Tests Shop Air 56 Tests Battery 60 charge 74 Tests Shop Air 51 Tests Battery 40 charge 56 Tests Shop Air 37 Tests Battery 20 cha...

Page 17: ...l 17 USA Contamination Standards See the Parker Guide to Contamination Standards DD0000015 available on your CD This publication makes available industry recognized cleanliness specifications for both...

Page 18: ...aning the icountBSplus touch screen If needed you may use a small amount of screen cleaning fluid sprayed directly onto the cloth and then applied to the touch screen If you need further advice please...

Page 19: ...re the power switch is in the OFF position 3 4 Check the plug is fully located Turn on power to charge battery Battery charging status on front panel and by flashing green LED on power switch See Batt...

Page 20: ...free cloth and cleaning materials If you experience difficulty contact Parker Hannifin Manufacturing UK Ltd for advice Connecting Disconnecting 1 2 Locate hoses Note that the colour of the hoses supp...

Page 21: ...power light may still flash if connected to a mains supply To remove the VAPOR RELEASE 4mm hose press down the connector sleeve and pull the hose out There may be a small fluid leakage here 3 4 To rem...

Page 22: ...ACC6NW005 carefully If you experience difficulty contact Parker Hannifin Manufacturing UK Ltd for advice 1 2 End of roll change paper Remove the printer cover 3 4 Remove the empty roller Insert empty...

Page 23: ...ountBSplus Bottle Sampler Manual 23 USA 9 bk Close the printer clip The paper is shown correctly loaded bl bm Feed paper through the cover Locate the cover in icountBSplus unit bn click Close the cove...

Page 24: ...ems on the touch screen as this may damage the product and is not covered by any warranty If in doubt please refer to the operator manual or contact Parker Hannifin Manufacturing UK Ltd for advice See...

Page 25: ...icator shows a possible fault condition NOTE If a fault indicator is illuminated please contact Parker Hannifin Manufacturing UK Ltd for advice 1 From the System Settings screen select Language to rec...

Page 26: ...sly sedimented contaminant being retained into the flowing oil It is also possible that these changes may cause partially contaminated filter elements to shed particulates into the system Samples shou...

Page 27: ...avoid distortion of the result by prolonged use which could cause the breakdown of contaminants Bottle samples can be sufficiently stirred by swirling and tumbling by hand end over end Samples should...

Page 28: ...d 28 USA Testing Inserting and removing sample bottles 1 2 Pull to open door Ensure the door latch is clear of front 3 4 Chamber ready for sample Insert bottle holding drip tray in one hand bottle in...

Page 29: ...may vary in size so the drip tray has a target to indicate that the bottle should be located centrally around the drip tube The bottle should look like this 7 8 Bring door front to latch Push door ful...

Page 30: ...ease contact Parker Hannifin Manufacturing UK Ltd if you need advice 1 2 From the System Settings screen select User Config for the main Test Set up screen Click on Bottle Name text box for a list of...

Page 31: ...plus Bottle Sampler Manual 31 USA 5 6 Type Bottle Name Prefix press Enter Click on text box press Add 7 8 New bottle name prefix is added to the Prefix list Name order priority highlight the name and...

Page 32: ...ority bl bm Your new name is selected To delete a bottle name highlight the line and press Delete Tick boxes give the option to delete old test data as memory becomes full or to delete test data after...

Page 33: ...UK Ltd for advice See Electrical connections pages 18 19 See Fluid connections if applicable pages 20 21 See icountBSplus product setup page 24 See Preparation guidelines for testing pages 26 27 and...

Page 34: ...Parker Hannifin Manufacturing UK Ltd 34 USA 3 4 Select Standard Set Flush Volume 5 6 Set Number of Samples Set Volume per Sample...

Page 35: ...e Sampler Manual 35 USA 7 8 Indicate if auto printing of results is required Set Start Test 9 bk Pressurising stage of testing Total progress progress of each test If you abort the test this warning s...

Page 36: ...screen Summary codes for bottle Results screen Total count values for bottle bn bo Results screen Count values for bottle with Print option Results screen If required view your results using another r...

Page 37: ...icountBSplus Bottle Sampler Manual 37 USA bp Results screen Browse all your tests...

Page 38: ...orting test results 1 2 Select Browse Tests from the main Test Set up screen List of Saved Tests is shown 3 4 Select individual results and show date You can double click the test name to view that te...

Page 39: ...the USB pen drive in the back of the icountBSplus USB memory used with the icountBSplus should have the following specification Flash memory USB 2 0 WIN USB powered 5V 7 8 See page 43 Importing Test r...

Page 40: ...Manufacturing UK Ltd 40 USA 9 CAUTION Data is lost once Delete button is clicked To delete results highlight the test result s for deletion using the stylus and select Delete A warning is displayed b...

Page 41: ...icountBSplus Bottle Sampler Manual 41 USA Test results Filtering 1 2 Browse results select Filter Filtering Options page 3 4 Sort Results by Date order Select the order of sorting for showing results...

Page 42: ...Parker Hannifin Manufacturing UK Ltd 42 USA 5 6 Filter Results by Date within a date range Filter Results by Prefix name 7 8...

Page 43: ...are other options for spreadsheet software Please contact Parker Hannifin Manufacturing UK Ltd for advice 1 2 Ensure the USB pen drive contains exported icountBSplus data Plug the USB pen drive into...

Page 44: ...Parker Hannifin Manufacturing UK Ltd 44 USA 6 The test results from the icountBSplus have been imported into the spreadsheet...

Page 45: ...rip tray 4mm 6mm blanking plugs Waste bottle 2x 250ml sample bottle pack Drain Vapor tube pack Push fit quick release air line adapter Documentation pack on CD Calibration certificate USB pen drive Sp...

Page 46: ...iption Part number Power pack UK 2m cable ACC6NW023 Power pack US 2m cable ACC6NW024 Power pack EUR 2m cable ACC6NW025 250ml Sample bottle kit x2 2 Bottles ACC6NW001 250ml Sample bottle kit x50 50 Bot...

Page 47: ...kit x5 ACC6NW016 Parker Transit Packaging Case ACC6NW020 icountBS2 manual CD ACC6NW026 Drip tray ACC6NW028 Alternative packaging single use ACC6NW030 Stylus pen ACC6NW031 Parker USB pen drive Flash me...

Page 48: ...Parker Hannifin Manufacturing UK Ltd 48 USA Description Part number Blanking plugs 4mm and 6mm ACC6NN033 Verification Fluid SERMISC067...

Page 49: ...Open door and remove door seal Use isopropanol to clean the inside of the channel and the inside of the door removing all traces of silicone grease Using a syringe apply silicone grease compound to t...

Page 50: ...ll amount of silicone grease to a gloved thumb and feed the O ring in evenly Coat the whole of the O ring evenly and then close the door Using the torch inspect the inner face of the door to ensure th...

Page 51: ...centers NOTE Return the IcountBSplus in its original Transit Packaging Case for any service calibration or warranty claims If the case is not available you can order suitable single use packaging see...

Page 52: ...Parker Hannifin Manufacturing UK Ltd 52 USA Troubleshooting icountBSplus is not emptying after a test icountBSplus reading zero counts after test Y Y N N...

Page 53: ...icountBSplus Bottle Sampler Manual 53 USA Y Y Y Y N N N N Diptube does not stay down...

Page 54: ...Parker Hannifin Manufacturing UK Ltd 54 USA Y Y Y Y N N N N Diptube stays down...

Page 55: ...icountBSplus Bottle Sampler Manual 55 USA Y N Y N Diptube operation not smooth Diptube stuck...

Page 56: ...Parker Hannifin Manufacturing UK Ltd 56 USA Y Y Y N N N icountBSplus not switching on power fault using power supply...

Page 57: ...icountBSplus Bottle Sampler Manual 57 USA icountBSplus not switching on power fault using battery Y Y N N...

Page 58: ...Parker Hannifin Manufacturing UK Ltd 58 USA icountBSplus gives suspect counts Y Y Y Y N N N N...

Page 59: ...icountBSplus Bottle Sampler Manual 59 USA Y Y Y N N N USB memory device not found upgrade failure...

Page 60: ...Parker Hannifin Manufacturing UK Ltd 60 USA Y Y N N Touchscreen failure...

Page 61: ...IE Ireland Dublin Tel 353 0 1 466 6370 parker ireland parker com IN India Mumbai Tel 91 22 6513 7081 85 IT Italy Corsico MI Tel 39 02 45 19 21 parker italy parker com JP Japan Fujisawa Tel 81 4 6635...

Reviews:

Related manuals for icountBSplus