background image

 
 
 

Section 1    Installation Instructions 

 

1.1 

Description 

These instructions describe the installation of the Digital 
MDM® mixer and a bag tee in a cabinet’s vertical 
surface. This manual includes installation instructions 
for an optional Matrx by Parker Scavenging System, as 
well as an optional gas outlet station. Figure 1 illustrates 
a typical installation. 
 

 

Figure 1. Outside View of a Typical MDM® 
Digital Cabinet Mount with Flush Mounted Bag 
Tee/Scavenging Control Valve 

 

IMPORTANT 

A bag tee containing an air inlet valve is required 
for use with this unit. The bag tee is provided 
separately for remote mounting. A standard 
Scavenging Nasal Inhaler or a set of rubber goods 
provides the patient connection. 
 
To reduce trac

e N2O in the operator’s environment, 

it is strongly recommended, though not required, 
that a Scavenging System be included with this 
installation. The scavenging system removes 
unwanted gases from the exhalation of the patient 
and the room. 

 

1.2 

Planning 

Planning is essential to the proper installation of the 
equipment. Without proper planning, minor changes 
can quickly become major problems. Read the entire 
installation instructions before attempting any and every 
installation. 
 
 
 

WARNING 

These instructions fully comply with the National 
Fire Protection Association (NFPA) guideline No. 
99C; see the Introduction for details. All equipment 
must be installed in accordance with these 
guidelines, as well as all local codes and 
ordinances

 

Matrx by Parker makes no warranty for cabinets 
and equipment that have been installed improperly. 

 

Planning should include the following: 

 

1.  The cabinet must be manufactured with the proper 

materials. 

 

  The mounting brackets are designed for 

mounting in wood, and wood and laminate, 
approximately 

3/4” thick. 

 

  Due to the nature of the mounting and the need 

for equipment stability during use, plaster board 
and other soft wall materials are not suitable. 

 

2.  Determine the location of the mixer head. 

 

  Typically, the mixer head is located at the 

approximate eye level of a seated person. 

 

  Consult with the practitioner to determine the 

desired location. 

 

3.  Determine the location of the bag tee. 
 

  The bag tee is either flush mounted in the 

cabinet surface (hoses are not externally 
visible), surface mounted to the cabinet surface, 
or mounted to a bracket inside the cabinet. 

 

  If a scavenging system is to be installed, it is 

mounted next to the bag tee. 

 

  The bag tee/scavenging control valve is supplied 

assembled to the appropriate mounting 
hardware. 

 

  Consult with the practitioner to determine the 

desired location. 

 
4.  If necessary, determine the location of the gas 

outlet station.  

 

  If the outlet station delivers only oxygen, it can 

be located next to the mixer head. If this is the 
case, the oxygen outlet is supplied assembled to 
the mixer head bezel. 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...Ref 10550800 Rev F Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA Tel 215 723 4000...

Page 2: ...g addresses Porter Regulatory Affairs EU Authorized Representative Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division 245 Township Line Road Riverside...

Page 3: ...ility A licensed electrical person should be consulted to confirm the Protective Earth ground connection Do not use an electrical cord with nicks cuts or other damage Do not use the unit with an exten...

Page 4: ...enging Control Valve Installation 6 1 6 J Hook Installation 8 1 7 Hose Connection 8 1 8 Rubber Goods Installation 9 1 9 Installation of Power Cord and Protective Earth ground Connection 10 Section 2 S...

Page 5: ...ns 5 Figure 4 Screw and Tee Nut Adjustment 5 Figure 5 Digital MDM Mixer Head Installation 6 Figure 6 Flush Mounting the Bag Tee and Scavenging Control Valve 6 Figure 7 Surface Mounting the Bag Tee and...

Page 6: ...tact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes Contact Information Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA...

Page 7: ...he fittings and tubing incorporate unique tubing sizes for each gas an important safety feature to prevent accidental crossing of lines during installation or maintenance The following basic requireme...

Page 8: ...information on the NFPA 99C code requirements consult your local Matrx by Parker representative or your local dental dealer U S Gas Color Coding All products in this manual are manufactured according...

Page 9: ...A guideline No 99C see the Introduction for details All equipment must be installed in accordance with these guidelines as well as all local codes and ordinances Matrx by Parker makes no warranty for...

Page 10: ...tion 1 Install the necessary gas plumbing from the central gas supply 2 If separate gas outlet stations are to be installed install the necessary gas plumbing See the Outlet Stations Installation and...

Page 11: ...4 mounting holes See Figure 4 6 While supporting the front of the cabinet tap in the tee nuts with a hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certa...

Page 12: ...installed The attachment to the cabinet surface must be sufficiently rigid to withstand the stress The bag tee scavenging control valve assembly may be installed on plywood attached to studs or on a...

Page 13: ...hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the...

Page 14: ...ace steps 8 9 and then installing the bag tee scavenging control valve to the mounting bracket steps 4 7 1 6 J Hook Installation The Digital MDM Flush Cabinet Mount Installation Kit includes a J hook...

Page 15: ...9 Figure 9 Connections Schematic some components may not be part of your installation Figure 10 Installation of Rubber Goods 1 8 Rubber Goods Installation Install the rubber goods See Figure 10...

Page 16: ...cord at the rear of the mixer head See Figure 11 Connect the power cord to the electrical outlet WARNING For safety a Protective Earth ground connection must be made through power cord 3rd wire or Gro...

Page 17: ...d the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head If the gas pipeline t...

Page 18: ...agent representative or dealer of Matrx by Parker shall constitute a warranty by Matrx by Parker or give rise to any liability or obligation of Matrx by Parker Subject to the next sentences Matrx by P...

Page 19: ...9600 Vacuum gauge scavenger control valve 91525136 Swing arm assembly with mounting hardware 91505112 Pre installation kit shutoff valves 91316043 3 liter breathing bag Mounting Fasteners 91525141 Har...

Page 20: ...ter Indexed Safety System 92305110 DISS female X DISS female Oxygen 92305111 DISS female X DISS female Nitrous oxide 92305248 DISS female X DISS female Vacuum DISS Hand I Twist 92305288 DISS Hand I Tw...