What happens
Alarm description
Default
Oxygen level too high
O2 high
Off
Oxygen level too low
O2 low
Off
Inlet pressure too high
P-inlet high
Off
Inlet pressure too low
P-inlet low
Off
Outlet pressure too high
P-outlet high
Off
Outlet pressure too low
P-outlet low
Off
Inlet temperature too high
T-inlet high
On
Inlet temperature too low
T-inlet low
On
Membrane pressure sensor fails
P-mem sensor fail
On
Outlet pressure sensor fails
P-Outlet sensor fail
On
Inlet temperature sensor fails
T-Inlet sensor fail
On
Status of temperature of compressor box 1
Temp comp1
On
Status of temperature of compressor box 2
Temp comp2
On
Oxygen sensor needs to be calibrated
Calibrate O2 cel
On
When an alarm is displayed there are two options:
1. Accept
2. Reset
When
ACCEPT
is touched, the alarm sound will disappear while the alarm level is still exceeded. If the alarm is not
resolved the alarm message will appear again in 24 hours. This function gives you some time to work on the
solution.
When
RESET
is touched the alarm status is cleared. However, if the alarm still exist it will appear again after the
delay time that has been entered in the alarm settings menu
(refer to § 6.3.2) has passed.