Warnings  and Precautions

 

These warnings and precautions are to help you to understand how to safely operate a 2-Cylinder 
Mobile Cart with a nitrous oxide & oxygen delivery System.  A WARNING alerts you to a possible 
hazard to people.  A CAUTION alerts you to the possibility of equipment damage. 

WARNING: Protect against crossed lines at the 2 Cylinder Mobile Cart regulator assembly.  Verify 

gas indexing pin locations, color coding of gauges, seal washer, & orientation of 
regulator assembly.  Follow instructions in Regulator Check page 6. 

 
WARNING: New or modified installations 

– properly connected gas pipelines (external gas supply) 

are absolutely essential to patient safety.  The ultimate responsibility of assuring that 
lines are not crossed rests with the user. 

 

WARNING:  Do  not  allow  crossed  lines  to  defeat  the  safety  features

  of  the  Mobile 

Cart, with any external gas supply, and/or central gas supply manifold systems.  Crossed 
lines  will  create  a  dangerous  and  hazardous  condition  where,  under  a  loss  of  oxygen 
supply,  100%  nitrous  oxide  will  be  delivered  through  the  oxygen  delivery  path  and 
subsequently to the patient.   
 
Use patient observation to prevent over sedation.  If a patient becomes over sedated when 
being  delivered  100%  oxygen  [during  an  apparent  loss  of  nitrous  oxide  supply],  it  is  a 
definite  indication  of  crossed  lines.    If  crossed  lines  are  suspected,  remove  the  mask 
immediately  and  encourage  mouth  breathing.    Deliver  pure  oxygen  from  an  oxygen 
demand valve only if the oxygen source is independent from the suspected crossed lines 
area.   

 
WARNING: 

Porter  Instrument  equipment  utilizes  the 

cross+protection

  system.  The  flexible  hose  and 

connectors that connect to the Regulator Assembly and the Flowmeter or Nitronox inputs are 
diameter indexed; 3/8” O.D. for Nitrous Oxide and ½” O.D. for Oxygen. The 

cross+protection

 

system is designed to prevent misconnection of Oxygen and Nitrous Oxide hoses.  

DO NOT 

ATTEMPT  TO  CHANGE  THE  DIAMETERS  OR  CONNECTORS  OF  THE  REGULATOR 
ASSEMBLY!

  Tampering with the 

cross+protection

 system constitutes acceptance of liability 

by the installer. 

 

WARNING: 

The Mobile Cart within a nitrous oxide & oxygen delivery system, with Flowmeter or Nitronox, 

is intended to be used by medical personnel trained in the device use and the use of nitrous 
oxide and oxygen for medical applications.  Do not use for other intended uses. Use approved 
Porter Flowmeter or Nitronox.     

 

WARNING: 

Do not use the Flowmeter or Nitronox device for the administration of general anesthesia or as 

a part of, or in conjunction with, a general anesthesia administration system.   

 

WARNING: 

Inspect  and  maintain  the  analgesia  delivery  system  to  prevent  N2O  leaks  in  all  hoses, 

connections and fittings. This includes the delivery system (Mobile Cart is mounting accessory) 
and  the  hoses  and  connections  to  the  Regulator  Assembly.  Have  all  leaks  repaired 
immediately. 

 

WARNING:

  Use  Scavenging:  Medical  workers  are  exposed  to  N

2

O  during  administration  of  N

2

O/O

2.

  

Controls  are  effective  in  the  patient  treatment  area  to  achieve  low  levels  of  ppm  (parts  per 
million)  exposure.    Controls  include  System  Maintenance,  Ventilation  and  Work  Practices.  
Your accessory Scavenger System is an important part of the system of controls in medical 
settings. 

Summary of Contents for Porter Nitronox

Page 1: ...ry Technique Safety Checks Maintenance and Troubleshooting Proper Use of Cylinders Connection to Flow Devices Use FM 1345 Assembly Instructions Examination of Contents Examine shipping carton for sign...

Page 2: ...e 1 5 Basket 1 6 Basket Clips 2 7 Cylinder Holder 1 8 Set Screw 1 9 Cylinder Wrench w Chain 1 Item Component Description Qty Image 10 Mounting Adapter 1 11 End Cap w 3 Screws 1 12 Hex Nut Adapter 1 13...

Page 3: ...3...

Page 4: ...rom the suspected crossed lines area WARNING Porter Instrument equipment utilizes the cross protection system The flexible hose and connectors that connect to the Regulator Assembly and the Flowmeter...

Page 5: ...rs A Loaded Mobile Cart must return to vertical when tipped 10 Do not roll cart aggressively For hardware protection follow Regulator Assembly configuration and location of nitrous oxide cylinder with...

Page 6: ...4 2 Check the color of the regulator gauges to ensure the O2 gauge is green white for ISO model the N2O gauge is blue 3 Check the indexing pins on each regulator to ensure they re indexed for the cor...

Page 7: ...m 12 screwed into the device mounting feature See Fig 7 illustration for mounting feature of Nitronox 2 Tighten the Hex Nut Adapter to lock the Mounting Adapter into place 3 Place the Mounting Adapter...

Page 8: ...Assembly see Fig 11 5 Slide the Regulator Assembly into place by lining up the pins and pin holes on the cylinder Push into place Properly placed the Regulator Assembly should be supported on the pin...

Page 9: ...adequate degree of fill before starting any procedure 5 Label cylinders with a tag or sticker indicating In Use and Full Full is reserve if appropriate 6 Use Cylinder Wrench to open the In Use cylind...

Page 10: ...he Utility Basket Note If location dimensions of the Utility Basket and Cylinder Holder were not strictly followed see FM 1345 2 Cylinder Mobile Cart Assembly Instructions the Scavenger Interface may...

Page 11: ...1 damaged or misused 2 repaired altered or modified by persons other than Parker Hannifin Corporation 3 not installed in strict compliance with applicable codes and ordinances or 4 not installed by a...