Part B: LabTec™ Methadone Software:
1.0 Software Overview: Main Menu
The LabTec has three (3) dispensing modes, namely, Serial, Volume and Weight, of which only the
Serial mode is used by clinics in normal use. This is a special version of the standard LabTec
designed for this particular application.
The Volume mode described is used only to attempt to free a stuck head as described later in the
Troubleshooting section.
The main menu of the LabTec Methadone consists of
four (4) operational modes
as shown on the
previous page. By using the
“Up”
and
“Down”
keys, one can readily scroll through the main menu.
Press the
“Select”
key to enter a chosen operational mode, i.e.
SERIAL.
By pressing the
“Select”
key the first level submenu is entered, which provides access to the
“Exec”
and
“Edit”
functions.
In the
“Edit”
submenu, the parameters for the application are selected.
SERIAL:
In the Serial mode, the LabTec is connected to PC running clinical software that sends
commands to the pump in the background. These commands are explained later in the manual.
VOLUME:
THIS MODE IS NOT USED OTHER THAN TO ATTEMPT TO FREE A STUCK HEAD.
CALL PARKER TECH SUPPORT IF YOU NEED TO IMPLEMENT THIS. IT WILL NOT REPLACE
THE SERIAL MODE IF THE COMPUTER IS DOWN.
WEIGHT:
THIS MODE IS NEVER USED IN THE CLINICAL SETTING.
MANUAL:
Allows manual pump speed control. This mode can be used to confirm the function of
the pump head, or for manual cleaning of the head.
SETUP:
This operational mode allows selection of various user preferences and interface options.
The unit comes properly configured and ready to use. Call Parker Tech Support of assistance with
Setup Mode is required.