background image

the 

Thermasonic® Gel Warmer

1. IMPORTANT SAFEGUARDS

This symbol              when indicated on the equipment means:

CAUTION:

Read these operating instructions fully before use and pay 

particular attention to sections containing this symbol.

Always observe the following safety precautions:

1.  Read all instructions.

2.  To provide continued protection against risk of electric shock,

connect only to a power supply that provides a safety ground terminal.

3.  Connect only to a power supply with a voltage corresponding  

to that on the electrical information label.

4.  Before moving, disconnect at the power supply socket.  

5.  If liquid is spilled inside the unit, disconnect it from the power 

    supply and have it checked by qualified personnel.

6.  Do not submerge the unit in water.

2. INCLUDED IN PACKAGE

3. ASSEMBLY

3.1 INSTALLATION

The Thermasonic Gel Warmer should be placed away from drafts, in
an upright position on a flat, stable surface. Do not lay the unit flat
on any side. The unit may also be wall mounted. See section 3.2
for wall-mounting instructions. Plug female end of power cord into
IEC320/C14 receptacle (see Figure 1) and male end into an 
appropriate electrical outlet matching the voltage listed on the 
electrical information label on the back of the gel warmer.

3.2 WALL MOUNTING

The Thermasonic Gel Warmer can be wall mounted with use of the 
supplied screws and wall anchors. Follow these steps to wall mount.

1. Once wall-mount location has been identified, make two marks on the wall 

1.75 inches (4.45 cm) apart. When making your marks, use a level to assure 
the gel warmer will appear straight when complete.

the 

Thermasonic® Gel Warmer

Warm and comfortable is now even better.

Operating Instructions

For Models: 82-01 (120V), 82-03-20 (230V) 

Thermasonic                      Operating                  Power Cord

Screws and

Gel Warmer                      Instructions

(IEC320/C14    

Wall Anchors

Compatible)

Figure 1

Warm and comfortable is now even better.

2.  If mounting in wood, install screws, and allow 0.1” (2.5mm) clearance

between the wall and the screw head. If mounting into drywall, the use of 
wall anchors is necessary.        

CAUTION:

Never drill or install screws in a

wall with wires or plumbing running through it.

3.  Carefully position the gel warmer so the large round wells (see  

Figure 2) on the back of the unit will pass over the screw heads.

4.  Press the gel warmer against the wall and slide downward, 

trapping the screw head in the grooves just above the wells.

4. OPERATION

4.1 HEAT INDICATOR LAMP

The Heat Indicator Lamp will blink while the unit is approaching 
the set temperature (see Figure 3). Once that temperature has 
been reached, the lamp will remain illuminated. Should the unit
restart the heating process, the Heat Indicator Lamp will again
begin to blink. 

4.2    OPERATING THE GEL WARMER

         

The gel warmer will warm a full 

bottle of gel to 104° F (40° C).

CAUTION: Temperature can vary. Always test a small quantity of gel before applying.

Allow approximately two hours for a full bottle of gel to reach proper temperature.
Your Thermasonic Gel Warmer is designed for continuous operation, and it is not
necessary to disconnect the warmer when not in use.

4.3    ERROR CODES

The Thermasonic Gel Warmer contains a temperature sensor to consistently
monitor functions and a Microtemp

®

thermal cutoff to protect against overheating.

If the temperature sensor is not operating at all, the Heat Indicator Lamp will blink
once every five seconds. If a short circuit occurs, the Heat Indicator Lamp will
blink twice every five seconds. In either circumstance, the heater will not operate.

5. TECHNICAL SPECIFICATIONS

This equipment is intended for indoor use and will meet its performance fig-
ures within the ambient temperature range of 68˚F (20˚C) to 95˚F (35˚C), with
maximum relative humidity of 80% (non-condensing). Installation Category II
(transient voltages). Pollution Degree 2 in accordance with IEC 664. Suitable for
operation at altitude of up to 6,500 feet.

Specifications:

Supply Voltage Range:      120V~ +/- 10%, 50/60 Hz

         

230V~ +/- 10%, 50/60 Hz (CE version)

Power Rating:                    Model 82-01: 12W, 0.1A

         

Model 82-01-20: 12W, 0.06A

 

104° F

 (40° C)

Figure 2

Figure 3

Therma_8201_Therma_single  2/4/13  5:48 PM  Page 2