Please read and understand this document before installing the filter housing and accessories.  Refer also to the 
general safety information in Table 2. 

 

 

 

Avoid subjecting the pressure equipment to external stresses. 

 

Do not use the filter housing or accessories if there are signs of damage either to the 
shell or sealing faces. 

 

 

Ensure that connecting pipework is adequately supported, so as not to impose local 
loads and vibrations onto the filter housing nozzles. 

 

Check for any damage on receipt of the product; inform the supplier immediately of 
any damage and await instructions whether to proceed. 

 

Packaged filter housings should be stored in a dry environment. 

 

Ensure the use of appropriate PPE during installation. 

 

Ensure necessary and correct handling, assembly, tightening and checking of all joints 

following installation and maintenance of the filter housing. 

 

Ensure that no stray electrical currents are being transmitted to the filter housing that 
may result in corrosion or perforation of metal parts; eliminate the source and /or 
earth appropriately. 

 

Perform a full system check before first use. 

 

 

Filter housings may be heavy – follow site safety controls for manual or mechanical 
handling to avoid harm. 

Table 5 

5.1

 

Receiving and Handling 

a)

 

Check for any damage on receipt of the product; inform the supplier immediately of any damage and 
await instructions whether to proceed. 

b)

 

It is recommended to unpack the filter housing near to final installation location to avoid unnecessary 
movement.   

c)

 

Use care when unpacking the filter housing to prevent damage to the contents. 

d)

 

Check that no small parts, documentation or certification are disposed of with the packing. 

e)

 

Filter housings may be heavy – follow site safety controls for manual or mechanical handling to avoid 
harm. 

f)

 

Immediately notify the supplier if there are too many parts or if parts are missing and await instructions. 

 

5.2

 

Storage 

a)

 

Packaged filter housings should be stored in a dry environment.   

b)

 

Use caution if stacking boxes and/or crates; always follow the instructions on the outside of packaging.   

 

5.3

 

Installation Location 

a)

 

Protect the installed equipment from mechanical damage that may impact the integrity of the pressure 
envelope (i.e. physical barriers and appropriate warning signs). 

b)

 

Ensure to check that the housing is not subject to vibration, thermal expansion and undue load on the 

nozzles or connections.   

c)

 

Use appropriate supports to prevent stresses and moments on the filter housing nozzles. 

d)

 

Ensure the filter housing is installed in a location with adequate clearance to allow cartridge change 

out. 

e)

 

The filter housing should be installed so that accidental leaks cannot be dispersed into the environment. 

f)

 

Take all due measures to prevent fires that could impact the integrity of the pressure envelope. 

 

 

Summary of Contents for VBA-11A

Page 1: ...179550500 Rev 2 Instructions for Use VBA Range...

Page 2: ...al Layout 6 4 3 Spare Parts 7 4 4 Data Label 7 4 5 Declaration of Conformity 7 5 INSTALLATION 8 5 1 Receiving and Handling 8 5 2 Storage 8 5 3 Installation Location 8 5 4 Connection and Assembly 9 6 O...

Page 3: ...authorized distributors provide product or system options for further investigation by users having technical expertise The user through its own analysis and testing is solely responsible for making...

Page 4: ...p to the coupling point connecting it to other equipment a vessel may be composed of more than one chamber Filter Cartridge Filter element installed inside the housing with a sealing feature PED Press...

Page 5: ...se spare parts that are not specific to the model or type of filter housing installed Compliance to National and local codes of practice environmental regulations and Health Safety directives have pre...

Page 6: ...ury or death Table 3 4 1 Product Description Refer to the Parker general arrangement drawing number 179503030 for details of the filter housing further information can be found in the product specific...

Page 7: ...d to the main filter housing body to provide important information about the specification see Figure 2 Pos Description 1 Parker Part Code 2 Design Temperature C 3 Design Pressure Bar 4 PED Fluid Grou...

Page 8: ...t of the product inform the supplier immediately of any damage and await instructions whether to proceed b It is recommended to unpack the filter housing near to final installation location to avoid u...

Page 9: ...all drain valves and discharges points to prevent fluid from harming personnel property or the local environment 6 Operation and Maintenance Please read and understand this document before operating o...

Page 10: ...ep 2 For disposal of the used filter cartridge refer to section Decommissioning and Disposal d Fit the replacement filter cartridge in accordance with the instructions provided in the filter cartridge...

Page 11: ...ondition which could lead to the onset of a hazardous situation d Replace worn parts with those designed for the filter housing Check the condition of filter housing seals each time the cartidges are...

Page 12: ...eases to work properly Replace closure locking ring or housing contact your local Parker representative Leakage of fluids from access connections to the tank Inadequately tightened connections Contami...

Page 13: ...0500 Rev 2 Page 13 of 13 www parker com bioscience Parker Hannifin Manufacturing Ltd Bioscience Filtration EMEA Durham Road Co Durham DH3 2SF England phone 44 0 191 4105121 email bioscience emea parke...

Reviews: