РА1.007.001РЭ-001 

16 

PARMA Ltd.

 

 

 

Figure 7 

6.2.5  Measurement of current-current phase shift 

6.2.5.1 

To  measure  current-current  phase  shift  (only  for  the  de-

vices  completed  with  two  clamps),  connect  reference  clamp  (see  fig.  8)  to 
connector  marked  with  Ireference.    Join  measuring  clamp  to  connector 
marked with Imeasuring. Use them for their designated purpose, as they are 
not  interchangeable.  Clamps  are  marked  for  correct  orientation  related  to 
source  (generator)  of  current.  In  case  of  false  orientation  additional  phase 
shift will be measured. 

6.2.5.2 

Embrace  the  wires  with  a  clamp  and  make  sure  that  it  is 

fixed well. Since the moment of signal appearance from clamp of the refer-
ence  channel  (Ireference)  the  device  will  be  switched  to  the  proper  mode 
automatically.  The  lower  line  will  display  the  value  of  phase  shift  between 
Ireference and Imeasuring. Read the indication from the lower line of display, 

prefix of current-current phase shift: 

ii=; 

 

Figure 8 

6.2.6  Measurement of current-voltage phase shift 

6.2.6.1 

To measure current-voltage phase shift, connect reference 

clamp to connector marked with Ireference (see fig. 9).  

6.2.6.2 

Apply voltage to terminals, marked as Uref. Since the mo-

ment of signal appearance from clamp of the reference channel (Ireference) 
the device will be switched to the proper mode automatically. The upper line 

─190.0W     

220.0V 

    29.2var     999 mА 

180° 

270° 

0° 

90° 

Uc 

Ia 

150° 

φiu=  ?° 

   0.0V 

φii= 30° С     999 mА 

180° 

270° 

0° 

90° 

Ic 

Ia 

30° 

РА1.007.001РЭ-001 

13 

PARMA Ltd.

 

4.7.2.2 

The device consists of two galvanically separated parts: 

reference and measurement channels. 

4.7.2.3 

The reference channel is intended for formation of signals 

necessary for measurement of phase shifts of galvanically isolated circuits. 

These may be current (Im) or voltage (Um). The reference channel consists 

of three comparators (1,2,3); two of them to form reference signals Im and 

Um (1,2) and one (3) to work with voltage signal Ub in the mode of determi-

nation of phase alternation order. Comparators convert the input sinusoid 

signal to meander and pass it through opto-isolators (9) to micro-controller 

(10) for further processing. Comparators’ supply is provided by isolated volt-

age converter (4). 

4.7.2.4 

The measurement channel apart from similar signal forma-

tion facilities for phase shift determination (7,8) contains two voltage-to-

frequency converters (5,6) used for direct measurement of current and volt-

age values. 

4.7.2.5 

All formed signals come to micro-controller (10), where 

programmed evaluation and operation mode choice is made based on de-

termined priorities (see below). The results of measurements are taken out to 

display (11). 

4.7.2.6 

The display has lighting controlled by timer (12) set on 10 

sec, which can be started by a button (13). 

4.7.2.7 

Power supply of measurement channel is provided by the 

device (14), forming voltage b5 V and –5 V. 

PREPARATION FOR USAGE 

5.1 

For device power supply use 4 galvanic elements of the type, 

mentioned in p. 4.3 with total voltage of 6 V. Use of analogues is permissible. 
To install power supply battery unscrew the screws on the back side of the 
device  and  install  4  elements  in  accordance  with  marking  on  the  cover  of 
battery section. 

5.2 

The device maintains serviceability in case of deep discharge 

of the battery without sufficient increase of measurement error. The feature 
of  necessity  to  change  a  battery  is  a  sharp  decrease  of  display  contrast 
range or its flicker. 

Summary of Contents for VAF-A

Page 1: ...ё РА1 007 001РЭ 001 PARMA Ltd Saint Petersburg VOLTAMPEREPHASOMETER Parma ВАФ А OPERATING MANUAL РА1 007 001РЭ ФГУ ...

Page 2: ...РА1 007 001РЭ 001 27 PARMA Ltd ...

Page 3: ...РА1 007 001РЭ 001 26 PARMA Ltd РА1 007 001РЭ 001 3 PARMA Ltd Appearance of Parma VAF A voltamperephasometer with its accessories ...

Page 4: ...ntal factors 8 5 Overall dimensions in packing are 330x235x105 mm 8 6 Gross weight is no less than 3 kg 8 7 The device is suitable for the use in the sphere of state metrological supervision in case when there are two unopened seals on its housing on the right and on the left belonging to the manufacturer enterprise and a Gosstandart body carrying out verification of the device 8 8 The device seal...

Page 5: ...or use without release verifica tion 7 8 Verification interval 7 8 1 Verification is made one time a year 8 MARKING PACKING AND SEALING 8 1 The following data is marked on a device name type manufac turer s trade mark national compatibility mark after type registration serial num ber year of release designation of input and output circuits nominal voltage type of current and power supply frequency...

Page 6: ...oltage current РА1 007 001РЭ 001 23 PARMA Ltd 7 6 4 4 VAF is considered to pass the test if the range of effective value of direct voltage is 0 0 460 V and the relative permissible error doesn t exceed the value determined in p 7 1 7 6 5 Verification of effective range and measurement error of effective value of alternating voltage measurement 7 6 5 1 Connect the device to UPPU 1M system 7 6 5 2 S...

Page 7: ...o value of voltage is shown on indicator 7 6 4 3 Make measurements in the following range points 10 100 200 300 460 V РА1 007 001РЭ 001 7 PARMA Ltd Um Im Measured value of voltage current 3 SAFETY REQUIREMENTS 3 1 The device belongs to class II according to GOST 26104 re garding electric injury protection 3 2 Protection level of the device case according to GOST 14254 80 IEC 529 is IP54 3 3 Only p...

Page 8: ...on linear distortion of supply line voltage no more than 15 Maximum consumed power no more than W From network power source 3 From battery source 0 5 Direct current power supply voltage is 6 V 4 galvanic cells R14P AR SIZE C 1 5 V SUM 2 Battery supply is switching automatically when network supply cable connector is removed from a jack РА1 007 001РЭ 001 21 PARMA Ltd Operation is made during Names ...

Page 9: ...thstand test voltage of effective value of 2 kV frequency of 50 Hz during 1 minute without damage 4 4 4 Electric insulation between network circuits and equivalent cir cuits and the case of the device can withstand test voltage of effective value of 4 kV frequency of 50 Hz during 1 minute without damage 4 5 Service requirements in overload conditions 4 5 1 The device can withstand overload of 2 Us...

Page 10: ...front panel 3 there is a power switch 5 display lighting button 6 double terminal 7 for connec tion of phase A voltage signal or Uref single terminal 8 for connection of phase B voltage signal double terminal 9 for connection of phase C voltage signal or Um jack 10 for connection of a clamp Iref jack 11 for connection of a clamp Im display 12 The jack for connection of the supply line of 220 V is ...

Page 11: ...ll be switched to required mode automatically when the signal comes to terminal B 6 2 9 2 Correct determination of phase shift order is possible if only all three phases are connected in accordance with marking on the de vice see fig 12 13 6 2 9 3 The result of phase shift determination is displayed as a text for example Figure 12 NO PHASE С NO PHASE А φuu 30 L 220 0V φui 30 C 999 mА 180 270 0 90 ...

Page 12: ...ent of signal appearance on Ureference terminals the device will be switched to the proper mode automatically 6 2 7 3 The upper line of display will indicate the value of phase shift between Ureference voltage and voltage applied to measurement chan nel Umeasuring Read indication from the upper line of the display 6 2 7 4 Prefix of current voltage phase shift uu Figure 10 6 2 8 Measurement of volt...

Page 13: ...is intended for formation of signals necessary for measurement of phase shifts of galvanically isolated circuits These may be current Im or voltage Um The reference channel consists of three comparators 1 2 3 two of them to form reference signals Im and Um 1 2 and one 3 to work with voltage signal Ub in the mode of determi nation of phase alternation order Comparators convert the input sinusoid si...

Page 14: ...А1 007 001РЭ 001 15 PARMA Ltd 6 2 2 Measurement of effective value of alternating voltage and its frequency 6 2 2 1 To measure the effective value of alternating voltage and its frequency apply alternating voltage to terminals named Um see fig 5 read display indication The measured voltage value is placed in the right upper part of the display and has a dimension of voltage V frequency value is in...

Reviews: