РА1.007.001РЭ-001 

22 

PARMA Ltd.

 

 

7.5 

Facilities for verification 

7.5.1  Megger F-4101, accuracy class 2.5. 
7.5.2  Universal  semi-automatic  test  system  UPPU-1M.  Voltage 

and current tuning range is 750 V; 10 A, error 0,03%. 

7.5.3  Voltmeter testing device V1-13. Range is up to 1000 V, er-

ror 32 mV. 

7.5.4  Phase  shift  meter  F2-34.  Phase  shift  measurement  range 

is 0 – 360 deg, error 0.01 deg. 

7.5.5  Note:  Other  types  of  measurement  facilities  are  permissi-

ble,  providing  stipulated  accuracy  of  measurement  and  verified  (cali-
brated) in stipulated order. 

7.6 

Verification conduction 

7.6.1  External examination 

7.6.1.1 

During  external  examination  it  is  necessary  to  make  sure 

that marking conforms to specifications requirements and there are no visible 
damages of case, power supply cord and plug, input commutation contacts, 
connector  assembly and  cable.  To  check  joint  of jaws  of  clamps  and make 
sure that the surfaces are clean. 

7.6.2  Determination of insulation resistance 

7.6.2.1 

Insulation resistance measurement is made with megger at 

a fixed voltage of 1000 V. 

7.6.2.2 

Measurement  is  made  between  input  contacts,  joint  to-

gether and contacts of power supply plug, joint together. 

7.6.2.3 

Insulation resistance is to be no less than 2.6 MW. 

7.6.3  Try-out 

7.6.3.1 

Connect the device to three-phase alternating current sys-

tem, three phase conductors are to be connected with red terminals A, B, C 
of the device. 

7.6.3.2 

Switch on the device. When shifting the connection of con-

ductors on terminals A, B, C  make sure that the display shows “Direct phase 
alternation” and “Reverse phase alternation”. 

7.6.4  Verification  of  effective range  and   measurement  error of 

direct voltage measurement 

7.6.4.1 

Connect the device to V1-13 system. 

7.6.4.2 

Switch on the device on and make sure that zero value of 

voltage is shown on indicator. 

7.6.4.3 

Make measurements in the following range points: 10, 100, 

200, 300, 460 V. 

РА1.007.001РЭ-001 

PARMA Ltd.

 

Um (Im)  - Measured value of voltage (current). 

SAFETY REQUIREMENTS 

3.1 

The  device  belongs  to  class  II  according  to GOST  26104  re-

garding electric injury protection. 

3.2 

Protection level of the device case according to GOST 14254-

80 (IEC 529) is IP54. 

3.3 

Only persons who had got safety instructions and studied the 

present Manual are permitted to work with the device. 

3.4 

It is prohibited to switch input circuits of the device with voltage 

in the researched circuits. 

3.5 

Switching on the researching circuits one should make all the 

operations with one hand. 

3.6 

Defining  the  phase  alternation  order  the  device  automatically 

connects  A,  B,  C  inputs  as  a  “star”,  the  resistance  between  any  two  “star” 
rays  being  2  MW.  Unused  terminals  (“cold”    in  phases  A  and  C)  are  com-
pletely disconnected. 

DESCRIPTION  OF  THE  DEVICE  AND  PRINCIPLES 

OF ITS OPERATION 

4.1 

Purpose 

4.1.1  Voltamperephasometer  “Parma  ВАФ-A”,  TU4221-006-

31920409-2004 is a completely automatic universal device. 

4.1.2  The  device  is intended  for measurement  of direct  voltage, 

effective value of voltage and current of sinusoidal alternating current with 
simultaneous calculation of active and reactive power in a circuit, meas-
urement  of  frequency,  voltage-voltage,  current-current  (if  the  device  is 
completed with two clamp), current-voltage phase shift and determination 
of phase alternation order in three-phase systems with midpoint or with-
out it. 

4.1.3  The  device  may  be  used  for  complex  protection  tests  of 

generators, transformers and lines, in current and voltage transformer cir-
cuits,  as  well  as  for  adjustment  of  phase-sensitive  relay  protection 
schemes and so on. 

4.1.4  The device is designed in accordance with requirements of 

GOST 22261 and current GSI (State System for Ensuring for Uniformity 
of Measurments) standards. 

4.1.5  The device OKP code is 42 21 99. 

Summary of Contents for VAF-A

Page 1: ...ё РА1 007 001РЭ 001 PARMA Ltd Saint Petersburg VOLTAMPEREPHASOMETER Parma ВАФ А OPERATING MANUAL РА1 007 001РЭ ФГУ ...

Page 2: ...РА1 007 001РЭ 001 27 PARMA Ltd ...

Page 3: ...РА1 007 001РЭ 001 26 PARMA Ltd РА1 007 001РЭ 001 3 PARMA Ltd Appearance of Parma VAF A voltamperephasometer with its accessories ...

Page 4: ...ntal factors 8 5 Overall dimensions in packing are 330x235x105 mm 8 6 Gross weight is no less than 3 kg 8 7 The device is suitable for the use in the sphere of state metrological supervision in case when there are two unopened seals on its housing on the right and on the left belonging to the manufacturer enterprise and a Gosstandart body carrying out verification of the device 8 8 The device seal...

Page 5: ...or use without release verifica tion 7 8 Verification interval 7 8 1 Verification is made one time a year 8 MARKING PACKING AND SEALING 8 1 The following data is marked on a device name type manufac turer s trade mark national compatibility mark after type registration serial num ber year of release designation of input and output circuits nominal voltage type of current and power supply frequency...

Page 6: ...oltage current РА1 007 001РЭ 001 23 PARMA Ltd 7 6 4 4 VAF is considered to pass the test if the range of effective value of direct voltage is 0 0 460 V and the relative permissible error doesn t exceed the value determined in p 7 1 7 6 5 Verification of effective range and measurement error of effective value of alternating voltage measurement 7 6 5 1 Connect the device to UPPU 1M system 7 6 5 2 S...

Page 7: ...o value of voltage is shown on indicator 7 6 4 3 Make measurements in the following range points 10 100 200 300 460 V РА1 007 001РЭ 001 7 PARMA Ltd Um Im Measured value of voltage current 3 SAFETY REQUIREMENTS 3 1 The device belongs to class II according to GOST 26104 re garding electric injury protection 3 2 Protection level of the device case according to GOST 14254 80 IEC 529 is IP54 3 3 Only p...

Page 8: ...on linear distortion of supply line voltage no more than 15 Maximum consumed power no more than W From network power source 3 From battery source 0 5 Direct current power supply voltage is 6 V 4 galvanic cells R14P AR SIZE C 1 5 V SUM 2 Battery supply is switching automatically when network supply cable connector is removed from a jack РА1 007 001РЭ 001 21 PARMA Ltd Operation is made during Names ...

Page 9: ...thstand test voltage of effective value of 2 kV frequency of 50 Hz during 1 minute without damage 4 4 4 Electric insulation between network circuits and equivalent cir cuits and the case of the device can withstand test voltage of effective value of 4 kV frequency of 50 Hz during 1 minute without damage 4 5 Service requirements in overload conditions 4 5 1 The device can withstand overload of 2 Us...

Page 10: ...front panel 3 there is a power switch 5 display lighting button 6 double terminal 7 for connec tion of phase A voltage signal or Uref single terminal 8 for connection of phase B voltage signal double terminal 9 for connection of phase C voltage signal or Um jack 10 for connection of a clamp Iref jack 11 for connection of a clamp Im display 12 The jack for connection of the supply line of 220 V is ...

Page 11: ...ll be switched to required mode automatically when the signal comes to terminal B 6 2 9 2 Correct determination of phase shift order is possible if only all three phases are connected in accordance with marking on the de vice see fig 12 13 6 2 9 3 The result of phase shift determination is displayed as a text for example Figure 12 NO PHASE С NO PHASE А φuu 30 L 220 0V φui 30 C 999 mА 180 270 0 90 ...

Page 12: ...ent of signal appearance on Ureference terminals the device will be switched to the proper mode automatically 6 2 7 3 The upper line of display will indicate the value of phase shift between Ureference voltage and voltage applied to measurement chan nel Umeasuring Read indication from the upper line of the display 6 2 7 4 Prefix of current voltage phase shift uu Figure 10 6 2 8 Measurement of volt...

Page 13: ...is intended for formation of signals necessary for measurement of phase shifts of galvanically isolated circuits These may be current Im or voltage Um The reference channel consists of three comparators 1 2 3 two of them to form reference signals Im and Um 1 2 and one 3 to work with voltage signal Ub in the mode of determi nation of phase alternation order Comparators convert the input sinusoid si...

Page 14: ...А1 007 001РЭ 001 15 PARMA Ltd 6 2 2 Measurement of effective value of alternating voltage and its frequency 6 2 2 1 To measure the effective value of alternating voltage and its frequency apply alternating voltage to terminals named Um see fig 5 read display indication The measured voltage value is placed in the right upper part of the display and has a dimension of voltage V frequency value is in...

Reviews: