4838 Reactor Controller

17

Datalogging and Charting

The 4838 Controller comes equipped with a charting 
and datalogging program.  This program cannot be 
run simultaneously with the control program, but 
once the user is satisfi ed that the controller is set, 
the control program may be closed, allowing the 
charting and datalogging program to run.  The mod-
ules on the 4838 may still be modifi ed manually at 
the physical panel during datalogging and charting.

1.  Open the

_

_

_

 

Com.exe fi le.  No installation is 

necessary.

2.  At the main program screen, select "protocol >> 

set PC" (if you have already done previously, you 
may skip this step).  Ensure that the following 
values are set: 

Baud:  9600
Data length: 8
Parity: even
Stop bit: 1
ASCII:  RTU

Click OK to return to the main screen.

3.  At the main screen, select "Program >> Record-

er".  The Recorder screen will come up.  

4.  Select the proper address for the Devices you 

wish to chart.  The primary temperature has ad-
dress 1.  A pressure display module will typically 
use address 3 and an HTM/ETLM will use ad-
dress 4.

5.  Click the Save button and save a REC 

fi le to an appropriate place on the PC.  
When the chart is generated, it will 
write to this fi le.

6.  When you return to the main Record-

er screen, notice that the Save button  
is depressed.  Hit the Start button to 
begin recording.  

7.  The chart will show the process values of each 

module represented.  As the values change, the 
chart will automatically scale to show the values.  

8.  When you wish to stop recording, click the stop 

button.  The chart has already been 
saved in the REC fi le created in step 5.  
You may view the chart by clicking the 
recorder button.

9.  You can convert the fi le to a TXT fi le for export to 

a spreadsheet as well.  

Converting a Chart File to TXT

Once a chart REC fi le has been created, you can con-
vert it to a TXT fi le for export into a spreadsheet.  

1.  Insert the 552M CD into the CD-ROM.  Locate the 

conversion fi le at “ \software\convert.exe”.  Copy 
it to the PC.

2.  Open the convert.exe fi le.  No installation is 

necessary.  

3.  Click the "CONVERT" button.  Locate the REC fi le 

to be converted, and click "Open".  

4.  The conversion program will create a new TXT 

fi le in the same location as the REC fi le with the 
name "CONVFILE.TXT".

Summary of Contents for 4838

Page 1: ...4838 Reactor Controller Operating Instruction Manual...

Page 2: ...ry Switches and Indicators 8 Protective Fuses and Relays 8 High Limit Reset 9 Access to the Controller Cabinet 9 Fuses 9 Instructions for the 4838 Reactor Controller Temperature Control Module 10 Comm...

Page 3: ...h of these controllers and describe the several optional control modules which can be added to these ex pandable units The users should study the instruc tions carefully before using any of these cont...

Page 4: ...4838 Controller There are no user serviceable parts inside the prod uct other than what is speci cally called out and discussed in this manual Advanced troubleshooting instructions beyond the scope of...

Page 5: ...l 4838 Controllers operate with aType J Iron Constantan thermocouple The total resistance of the thermo couple and the lead wires should not exceed 100 ohms If the resistance of the thermocouple circu...

Page 6: ...pera tion within the limits speci ed for that design But in order to preserve the validity of these de signs all temperature and pressure limits must be observed and no attempt should be made to incre...

Page 7: ...troller will not be affected by radiant heat Connections Labeled connections are provided on the rear panel of the controller Heating Only The heating output connector is to be used only with Company...

Page 8: ...When the light comes on the user must nd the source of the problem correct it and then manually reset the system using the high limit reset button on the front of the controller The Main Power Switch...

Page 9: ...imit setting This must be manually reset before heating can occur again On Model 4838 Controllers the high temperature limit is preset at the factory to 375 C for most sys tems On these controllers an...

Page 10: ...the value of the setpoint by adjust ing the heater output The setpoint may be adjusted by doing the following 1 Press the up or down arrow until the lower display shows the desired setpoint The lower...

Page 11: ...p and Soak The 4838 Controller has a programming feature which allows more exibility when compared to set point control Rather than controlling temperature to a single set point a ramp and soak progra...

Page 12: ...set key to solidify the change 11 Press set to return to the main screen Note that the lower display of the main screen has changed to show the program step Note The program will attempt to hold the...

Page 13: ...r knows the expected heat up time It is useful to keep this step 02 separate from step 01 so you can adjust step 01 to match the heat up time The time is zero making this a step function in stead of a...

Page 14: ...involving reactants corrosive to stainless steel the transducer body can be replaced with an isolator of special alloys which will keep the reactants away from the transducer Users should consult wit...

Page 15: ...As an added safety feature there is also a second ary alarm which will trip the High Limit Reset when breached This secondary alarm provides a redun dant high temperature safety cut off which will sh...

Page 16: ...The soft ware is located on the CD at software _ _ _ Com exe 6 Once copied to the PC open the _ _ _ Com exe le No installation is necessary 7 At the main _ _ _ Com program screen select protocol set P...

Page 17: ...se address 3 and an HTM ETLM will use ad dress 4 5 Click the Save button and save a REC le to an appropriate place on the PC When the chart is generated it will write to this le 6 When you return to t...

Page 18: ...Out1 read only Heater output If using a type J or type K thermocouple SP 0 If using an RTD SP 1 Press and hold down SET for 5 sec Operation Mode Select type value Comment InPt J Input type tPUn C Temp...

Page 19: ...ol mode ON OFF MANUAL PID and PID PROG S HC Heat Heat Cool control ALA1 2 Alarm operates when PV value is higher than SV Value ALH setting SALA OFF System alarm CoSH ON Communication write function th...

Page 20: ...nge tP L 0 Lower limit temperature range CTRL ON OFF Control mode ON OFF MANUAL PID and PID PROG S HC Heat Heat Cool control ALA1 6 Alarm operates when PV value is higher than AL1H value SALA OFF Syst...

Page 21: ...ontrol mode Select heating cooling control or dual loop output control Alarm 1 mode setting Alarm 2 mode setting Only available on PrimaryTemperature Controller Alarm 3 mode setting Only available on...

Page 22: ...function is enabled Lower limit alarm 1 This parameter is available only when ALA1 function is enabled Upper limit alarm 2 This parameter is available only when ALA2 function is enabled Lower limit a...

Page 23: ...control mode or Heating and Cooling control cycle Available in PID control mode Control cycle setting of 2nd output group Available in PID control and Dual Loop output control mode P value of 1st 2nd...

Page 24: ...tension wire type J 25 ft A470E6 Thermocouple extension wire type J 30 ft A470E9 Thermocouple extension wire type J 35 ft A470E10 Thermocouple extension wire type J 40 ft A472E4 Thermocouple type J 5...

Page 25: ...s 30 ft covered A1905E10 Pressure harness 35 ft covered A1905E11 Pressure harness 40 ft covered A1906EP02 Transducer 200 psi A1906EP05 Transducer 500 psi A1906EP10 Transducer 1000 psi A1906EP20 Transd...

Page 26: ...ft A470E10 Thermocouple extension wire 40 ft A472E Thermocouple type J 7 1 2 length A472E2 Thermocouple type J 9 1 2 length A472E3 Thermocouple type J 11 1 2 length A472E4 Thermocouple type J 5 1 2 le...

Reviews: