Page 8-10
Lasair II Particle Counter Operations Manual
Chapter 8: Statistics Mode
Statistics Setup Screens
The
Statistics Setup
screen is where you will make the settings for the standard you
will be using and the area you will be sampling. The
Statistics Setup
Screen has two
pages, the
Main
page and the
Setup
page.
To open the Statistics Setup Screen:
1
From the Main Display screen, press the Sampling Setup
icon
softkey.The Sampling Setup screen will open:
2
Press the
Statistics Setup
icon
softkey to open the
Statistics Setup Main
page.
Statistics Setup Main Screen
Name:
Enter the name of the area that will be certified. Example:
Cleanroom B
.
This name is included in the printed report CSV file output.
User:
The name of the person who will be collecting the data. This is printed in
the report.
Standard:
Select one of the standards or the
Average
function.
Class
Select the appropriate cleanroom class from the list.
Figure 8-1: Statistics Setup Screen - Main Page
Before the edit mode is activated, this icon’s softkey will toggle
between the
Main
and
Setup
pages.