Using Edit Mode - - General Procedure
Lasair II Particle Counter Operations Manual
Page 5-3
Using Edit Mode - - General Procedure
Setting parameter values in Lasair II particle counter Setup Screens
1
Verify that the screen has a blue pencil icon in the upper left
corner. If the icon is present, the screen is a setup screen and
values can be changed.
2
Press the Enter key to enable the Edit Mode.
3
Press the Tab Keys to move the cursor to the parameter to be
changed.
4
Press the Enter key to access the parameter.
5
Press the arrow or Enter keys to change the parameter.
6
Press Tab or Enter to record the desired setting.
7
Use the Return Arrow softkey to Exit the Edit Mode.