Configuring Communications
Lasair II Particle Counter Operations Manual
Page 7-3
Configuring Communications
All communications are configured using the “Comm Setup” screens on the Lasair
II particle counter. Figure 6-1 is an overview of the two communications setup
screens. One screen is used for all Ethernet based communications, and the other
screen is used for all serial communications. Each screen is discussed separately
below.
Ethernet Communications Setup Screen
Figure 7-2 is the Ethernet communications setup screen where all network
addressing parameters are set, and various network communication protocols are
enabled or disabled. When a network address is changed, the particle counter must
be rebooted. A pop-up message informs you of this when you try to exit the screen.
The various setup parameters are discussed below.
Figure 7-1: Setup paths for communications modes
Serial screen
Network screen
(Ethernet)
Lasair II
particle counter
communication
❏
Enable Web Server
(Select for web browser interface)
❏
Enable TCP/IP Remote Mode
(Select when operating in the
Remote Mode)
❏
Enable TCP/IP
(Select for PMS TCP/IP protocol
with Facility Net and Remote Mode)
❏
PMS Multi-Drop
(Select for MiniLaz emulation
with Facility Net)
❏
Lasair II particle counter
(Select for normal RS-232
communications)
COMM Setup Screens