background image

 

 

PAS V Sensitivity Checking 

 

 
 

NOTE

If  the  low  battery  lamp  starts  flashing  at  any  point  in  the  procedure, 

recharge the battery  before proceeding.

     

 
1.      Turn on the flashlight to activate the fuel cell heater.  Leave it on for 5 minutes so 

that the fuel cell comes up to design temperature before starting the test.     

 

2.      Check that the inlet and outlet ports of the PAS V are free of obstruction.     

 

3.      Hold the PAS V and insert the tip of the Calibration Assembly firmly into the PAS  

inlet  port.    Place  the  opposite  end  of  the  Calibration  Assembly  modified 
mouthpiece  against  the  simulator  outlet.      The  bar  graph  display  will  be  facing 
upwards.     

 

4.      Check  that 

the  simulator  is  running  at  34◦C  and  take  a  deep  breath  and  blow 

slowly  but  steadily  into  the  simulator.  (If  Cal-Pump  is  used,  follow  Cal-Pump 
instructions.)    After  5  seconds,  tap  the  Sensor  Control  Button  on  the  PAS  V  to 
start a sampling cycle. Continue to blow for another 5 seconds.     

 

5.      Look for the green sensor lamp and the yellow sampling lamp and listen for the 

sound  of  the  air  pump  which  should  come  on  immediately  after  the  Sensor 
Control  Button  is  activated.  Keep  blowing  until  the  PAS  V  pump  stops  (pump 
lamp goes out) 5 seconds after sampling began.     

 

6.      Disconnect the Calibration Assembly and observe the bar-graph display.  

All but 

one

 of the bars should light up. There should be two green, four yellow and two 

of the three red bars lit 15-20 seconds from the time the pump stops.     

 

7.      Record  the  number  of  red  bars  that  light  up  (this  should  be  two).  Repeat  the 

sensitivity test two more times noting each time how many red bars light up. If in 
two out of the three tests, two red bars are illuminated, the sensitivity of the unit 
is  satisfactory.  Note  that  the  number  of  red  bars  lit  may  vary  by  one  in  either 
direction  because  of  variations  in  the  rate  and  strength  of  blowing  through  the 
simulator.  This is the reason for conducting three tests.  If two out of the three 
tests do not result in the lighting of red bars (a blinking light counts), then the unit 
should be recalibrated following the calibration procedure described below.     

 

 
 

NOTE:

  

A minimum of 5 minutes should pass between each sensitivity test and an 

air blank should precede each sensitivity test.

     

 

 
Periodic functional checks are recommended.  These quality control  checks can 
be performed by simply holding the intake port near a known source of alcohol, 
activating  the  pump  and  observing  the  bar  graph  display  for  indications  that 
alcohol vapor is being detected. This will verify  the pump assembly is operating 
properly and the fuel cell sensor is functioning. 

 

Summary of Contents for PAS V

Page 1: ...Passive Alcohol Sensor CALIBRATION MANUAL PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com Copyright 2015 by PAS Systems International Inc All Rights Rese...

Page 2: ......

Page 3: ...Bath Solution optional 2 Operating the Alcohol Wet Bath Simulator 3 Readying the PAS V 5 PAS V Sensitivity Checking 6 PAS V Calibration 7 Dry Gas Method Equipment Checklist for Dry Gas Sensitivity Che...

Page 4: ...ece that connects the outlet of the wet bath simulator to the intake port of the PAS V Available from PAS Systems A NHTSA approved Wet Bath Simulator 500 milliliters of 005 ethanol alcohol solution To...

Page 5: ...volumetric flask and fill the flask up with distilled water This completes your mixture of stock solution You may keep your stock solution in a refrigerator for mixing calibration solution as necessar...

Page 6: ...jar from the simulator by unscrewing it from the top assembly Caution DO NOT remove Top Assembly and expose the heating element to open air with power turned ON This will result in damage to the heati...

Page 7: ...hts come on after observing for 15 20 seconds the air blank check is OK d If however one or more of the indicator lights come on again this clearance process should be repeated If after two attempts t...

Page 8: ...sampling began 6 Disconnect the Calibration Assembly and observe the bar graph display All but one of the bars should light up There should be two green four yellow and two of the three red bars lit 1...

Page 9: ...activate the Sensor Control Button on the PAS V 6 Look for the green sensor light and the yellow pump PMP light and listen for the sound of the air pump that should come on immediately after the Sens...

Page 10: ...ct the inlet outlet ports of the PAS V to make sure that they are not clogged 3 Conduct an air blank a Turn on the flashlight and wait a minimum of 1 2 minutes for initialization process to complete b...

Page 11: ...r blanks should be made away from the test room to see whether the unit gives a zero reading when removed from the test area 5 Once the air blank has been performed with a satisfactory result the PASV...

Page 12: ...d light up There should be two green four yellow and two of the three red bars lit 15 20 seconds from the time the pump stops 9 Record the number of red bars that light up this should be two Repeat th...

Page 13: ...ough the opening and into the calibration adjustment pot screw Turn the calibration adjustment all the way up counter clockwise The adjustment is free to rotate approximately 270 Take care not to use...

Page 14: ...Problem Reported Checked By Action Taken Date Returned to Manufacturer Date Returned to Field The example shown provides a means for recording any reports from Testers of problems with the PAS V units...

Page 15: ......

Page 16: ...PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Printed 10 15...

Reviews: