Wireless Drop Counter
Assembling the Titration Setup (See figure on next
8
013-16026A
10.
Step 5 opens automatically and it shows the “New
Calibration” in the lower part. NOTE: In this example,
Capstone counted 10 drops. Since the volume was 5.0
mL, the “New Calibration” shows “Slope” equal to
“0.500 mL/drops”.
11.
In Capstone, click “Finish” to return to Step 1. Click the
“Calibration” icon in the Tools panel to close the
Calibration window.
Calibration (Generic Process using SPARKvue on a
Computer or Tablet)
1.
Attach the Drop Dispenser (syringe, two stopcocks and
drop tip) to a support rod.
2.
Fill the syringe with titrant. Put a beaker under the Drop Dispenser to catch the titrant.
3.
Open both stopcock valves and adjust the valve on the bottom stopcock so that the drops fall at a rate of about
1 drop per second. Close the top valve after the rate is achieved. Discard the titrant as directed by your
instructor.
4.
Align the drop tip with the rectangular window in the Wireless Drop Counter. Put a graduated 10 mL cylinder
under the Drop Counter window so it can catch drops from the drop tip and give you a measure of the volume.
5.
Start SPARKvue.
•
The “choose a path” window opens.
6.
To “pair” the Wireless Drop Counter
wirelessly to the computing device
(e.g., computer or tablet), click or
touch “Sensor Data” in the “choose
a path” window.
•
The “Sensor Data Configuration”
window opens
.
7.
In the “Sensor Data Configuration
window”, look in the “Connected
Devices” list. Touch or click to select
the correct address that matches the
Device ID XXX-XXX number found
on the Wireless Drop Counter.
•
A message confirms that the sensor
is being connected.
Click or Touch
“Sensor Data”
Click or Touch