background image

®

PS-3517

Introduction

3

012-15719A

Software Help

See the SPARKvue Help or PASCO Capstone Help for 

information about collecting, displaying, and analyzing 

data. 

In SPARKvue, select the HELP button in any screen 

including the Home Screen.

In PASCO Capstone, select PASCO Capstone Help 

from the Help menu, or press F1.

Checking Electrode Operation (Slope)

1. Connect your electrode to the input connector on the 

Wireless pH /ISE/ORP Sensor (or other measuring 

device). 

2. Place 100 mL of DI water into a 150 mL beaker. Add 

1 mL Carbon Dioxide ISA to the DI water and stir 

throughly. Check the pH of the solution with the 

Wireless pH/ISE/ORP Sensor or a piece of lab pH 

paper. 

The pH value must be below pH 4.0. Otherwise, add 

Carbon Dioxide ISA in 1 mL increments until the pH 

value is below pH 4.0.

3. Rinse the electrode with DI water, blot dry, and 

place it in the beaker with the prepared solution.

To prevent air entrapment on the membrane 

surface, use an electrode holder that keeps the 

electrode at a 20 degree angle.

4. Begin recording. Add 1 mL of 1000 ppm Carbon 

Dioxide Standard Solution into the beaker. Stir thor

-

oughly, then record the potential (E

1

) in mV when a 

stable reading is displayed. 

5. Pipet 10 mL of the same standard solution into the 

beaker. Stir thoroughly. When a stable reading is 

displayed, record the potential (E

2

) in mV.

6. The difference between the first and second poten

-

tial readings, (E

2

-E

1

), is defined as the electrode 

slope. The normal range for the slope is 52 to 60 mV 

at 25°C.

Troubleshooting

If the electrode slope is not within the normal range 

stated above, or the electrode mV reading drifts, the 

Reference Fill Solution may need to be replenished. 

Hold the electrode body with one hand, and then gently 

pull the cable. Release the cable. Fresh fill solution will 

now occupy the space between the tip of the inner glass 

body and the membrane. If this procedure does not 

restore electrode response, follow the steps below:

1. Check the slope of the internal glass electrode. 

Remove the internal glass electrode from the 

electrode body.  Connect the internal glass 

electrode to the Wireless pH Sensor or to a pH/mV 

meter. Put the internal glass electrode in pH 7.00 

buffer. The annular ceramic junction should be 

immersed in the solution. Record the reading in mV 

when it is stable. Remove the internal glass 

electrode from pH 7.00 buffer. Rinse with DI water, 

blot dry and put it in pH 4.01 buffer. Record the 

reading in mV when it is stable. The reading in pH 

7.00 should be 0 + 25 mV. The millivolt difference in 

the two buffers should be greater than 168 mV. (If 

the difference is less than 160 mV, contact the man

-

ufacturer). 

2. If the internal glass electrode functions properly, but 

the completely assembled electrode does not 

function properly in the standards, replace the Elec

-

trode Body with a new one. Unscrew the Electrode 

Body from the cap. Get the extra Electrode Body 

from the plastic bag. Install the new Electrode Body. 

Follow the procedures described in “Electrode 

assembly” and “Checking electrode operation 

(slope)” again. 

If the electrode still does not operate properly, contact 

the manufacturer. 

3. Repeat "Checking Electrode Operation" procedure 

again. Note: All standard solutions should be 

prepared fresh. For best performance use ISA in all 

solutions.

Reading a Sample with the Electrode

Various procedures may be used to determine the 

concentration of a sample. The most common is the 

Direct Calibration method, which is described below. 

Contact PASCO's Technical Support department for 

details of other methods.

In Direct Calibration a series of standard solutions of 

differing concentrations is used to calibrate the 

electrode. Then each sample requires only a single 

measurement device reading, which is compared with 

the calibration readings to obtain the sample 

concentration. ISA is added to all solutions to ensure the 

samples and the standards have the same ionic 

strength.

Set up:

1. Prepare the electrode as described in "Electrode 

Assembly" and "Checking Electrode Operation". 

Summary of Contents for PS-3517

Page 1: ...d area Boot 1 Cover DO NOT let the BNC connector at the top of the unit get wet Included Equipment Item Included Equipment Item Carbon Dioxide ISE Electrode Body 2 1 CO2 Standard Solution 1000 ppm 4 Carbon Dioxide ISE Internal Electrode 2 CO2 Reference Fill Solution 5 BNC Male to Female Cable 3 CO2 Ionic Strength Adjuster ISA 6 ...

Page 2: ...e tip of the internal glass electrode Keep the boot for later use Note Soaking the internal glass electrode in internal filling solution or pH 4 01 buffer for at least 1 hour before proceeding with the following steps will ensure fast response 2 Remove an electrode body from the box Using the solution provided fill the electrode body with about 3 5 mL of Reference Fill Solution 3 Rinse the interna...

Page 3: ...ctrode response follow the steps below 1 Check the slope of the internal glass electrode Remove the internal glass electrode from the electrode body Connect the internal glass electrode to the Wireless pH Sensor or to a pH mV meter Put the internal glass electrode in pH 7 00 buffer The annular ceramic junction should be immersed in the solution Record the reading in mV when it is stable Remove the...

Page 4: ...100 mL of the sample into a 150 mL beaker Add 1 mL of ISA Stir thoroughly Check the pH value of the standard 6 Rinse electrode with Dl water blot dry and place in the sample beaker Wait for a stable reading and the sample concentration will be displayed on the mea suring device 7 Determine the sample concentration using the data analysis tools in the software Electrode Storage Short Term over nigh...

Page 5: ...ay be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulations that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws...

Reviews: