30
-
Conf: contains the indication of the first line relative to the source (energization
point), the second line refers to the acquisition of the trigger, and all the
following lines refer to the authorized geophones and are numbered in sequence
from 1, these numbers are those to which reference will be made later in the
process of noise monitoring, data acquisition display and saving on seg2 file.
-
Dev: identifies the instrument to which the geophone is connected using index 1
to identify the instrument with the lowest serial number, and index 2 to identify
the instrument with the highest serial number. The trigger should always be
connected to instrument 1.
-
Phy: identifies the physical geophone relative to the configured geophone, the
position number from 1 to 24 is reported on the line as shown in the figure of the
line configuration window.
-
X, Y, Z and S: parameters of the position on the axes x, y and z, and sensitivity
configured for each geophone.
For the Gea24 system with external PC, you can change the parameters directly on the
table, select the value with the mouse or with the tab button and enter the new value via
the keyboard.
For the Gea24 system with embedded PC you cannot change the parameter values
configured directly on the table. To change the configuration values you must use an
additional viewing window and change the parameters of a single geophone. To access
the edit window you have to press any cell within the display table. The edit window of
a single geophone is described in Section 6.3.4.
The window has the button "<AcqConfig", when pressed, it takes you to the acquisition
configuration window.
For the Gea24 system with external PC you will have the "All geophones" button that
allows you to open the parameter configuration window of all geophones
simultaneously. This window is described in Section 6.3.5.
6.3.4
Viewing and changing the configuration of a single geophone (only
for GEA24 embedded PC)
This window is only present on the Gea24 system with embedded PC, and is used when
entering the parameters of the geophones.
Figure 9 is an example of the window.
Summary of Contents for GEA24
Page 2: ......
Page 5: ...GEA24 SEISMOGRAPH 5 Table of Contents ...
Page 7: ...GEA24 SEISMOGRAPH 7 Important Notice ...
Page 9: ...GEA24 SEISMOGRAPH 9 Warranty and safety instructions ...
Page 11: ...GEA24 SEISMOGRAPH 11 Introduction ...
Page 54: ...54 Appendix A Technical Specifications ...
Page 55: ...GEA24 SEISMOGRAPH 55 Appendix B Appendix B How to place the piezoelectric shock sensor ...
Page 56: ...56 Appendix C Trigger connector FIGURE 17 TRIGGER CONNECTOR ...
Page 58: ...58 48 Channel seismograph 24 Channels cables ...
Page 59: ...GEA24 SEISMOGRAPH 59 Appendix D Line Connections ...