background image

6

Care and cleaning 

Seat unit and back rest

After bathing rinse the bath lift (without covers) with warm water. If the bath lift
is heavily soiled, wash it down with commercially available cleansing
detergents and brushes.

Remove the seat and backrest covers.

Dry the product with a cloth.

Use a mild disinfectant if required.

Never clean the drive or hand controller with a high pressure washer.

Whilst cleaning the seat unit be careful of the mobile lifting scissors.

Do not use strong solutions of disinfectant.

Ensure that the bath lift is completely dry before storing it.

We recommend cleaning the Bathmaster Deltis thoroughly when changing
location or patient.

Hand controller and battery charger

Please read the hand controller instructions.

Prior to cleaning, the charger must be removed from the power supply plug.

Never dip the charger into water, spray wash or wash with water.

Clean the charger with a cloth.

Clean the hand controller with a damp cloth and then dry off.

Ensure that the bath lift is completely dry before storing it.

Seat and backrest covers

Once removed from the main unit, the seat and backrest covers can be
washed at 40°C (max) in the washing machine.

Do not dry the covers in a tumble dryer.

Technical specifications (See Fig. 2)

Total weight without hand control:

Deltis

|

Deltis XL ..................10.5kg

|

11.1kg

Weight of seat unit:

Deltis

|

Deltis XL...............................................6.8kg

|

7.4kg

Weight of back rest

.........................................................................................3.7kg

Weight of hand controller:

Deltis

|

Deltis Premium & Deltis XL ..........0.3

|

0.4kg

Seat depth (A):

............................................................................................500mm

Max seat depth (B):

.....................................................................................455mm

Min seat height:

.............................................................................................65mm

Total depth (back rest upright) (C):

............................................................690mm

Total depth (back rest lowered) (C):

..........................................................900mm

Total height (D):

.........................................................................................1105mm

Height of back rest (E):

...............................................................................680mm

Width of back rest (F):

................................................................................350mm

Width of seat with side flaps (G):

...............................................................710mm

Width of seat without side flaps:

................................................................375mm

Footprint of the base plate:

..............................................................530 x 220mm

Backrest reclines to:

...........................................................................................40°

Load capacity:

Deltis

|

Deltis XL ....................................................140kg

|

170kg

Actuator degree of protection against water ingress:

...................................IP66

Hand controller degree of protection against water ingress:

.......................IP67

Power on time:

.................................................................10%/2 min - 90%/18 min

Supply voltage: 

Deltis 

|

Deltis Premium, Deltis XL & Deltis US ......11.1V

|

14.8V

Operating voltage: 

Deltis 

|

Deltis Premium, Deltis XL & Deltis US...........5A

|

5A

Rechargeable Battery:

....................................................................................Li-ion

Charge time when low voltage indicated: 

Deltis

|

Deltis Premium, 

Deltis XL & Deltis US .............................................................................5h

|

1h

Decibel Level:

.................................................................From/To 49db(A)/47db(A)

Charger: 

Deltis 

|

Deltis Premium & Deltis XL 

|

Deltis US

Input:

...............................................................................................230V AC/50Hz

..................................................................................

230 - 240V AC/50Hz

...............................................................................................

120V AC/60Hz

Output:

..................................................................................15V DC/500mA/7.5VA

................................................................................

20V DC/120mA/10VA

....................................................................................

20V DC/500mA/10VA

Warranty 

Patterson Medical Ltd warrants that the Bathmaster Deltis Bath Lift is free from
defects in material and workmanship. This warranty shall remain in effect for
three years from the date of original consumer purchase.

In the event that the unit needs inspection or repair, contact Patterson Medical
Ltd, the distributor or an approved agent.

This warranty does not cover the following:

Replacement parts or labour furnished by anyone other than Patterson
Medical Ltd, the distributor or an approved agent.

Any failure of the unit during the warranty period if the failure is not caused by
defect in material or workmanship or if the failure is caused by unreasonable use.

Patterson Medical Ltd is not liable for incidental or consequential injury or
damage to property.

Should it be necessary to return the Bathmaster Deltis, it is essential it is
adequately packed to protect it in transit, preferably in its own purpose made
original box, otherwise Patterson Medical Ltd cannot accept any responsibility
for transit damage to the unit.

Any attempt at repair or maintenance by persons other than Patterson Medical
Ltd trained or approved personnel will void this warranty.

Maintaining the Unit

The rechargeable battery in the hand control is not interchangeable and may
only be opened by the manufacturer.

The Bathmaster Deltis should be inspected before each use to ensure it is free
from damage and excess wear. The Bathmaster Deltis is built to give long,
reliable use, without need for servicing. Whilst being robust, the lift should be
treated with care and in particular should not be dropped. If damage or
deterioration is suspected, please contact Patterson Medical Ltd or the
distributor, for repair by an approved agent.

For professionals issuing the Bathmaster Deltis, Patterson Medical Ltd
recommend that it receives bi-annual inspections in line with requirements of the
Lifting Operations and Lifting Equipment Regulations 1998 (LOLER). Inspections
should only be carried out by trained and experienced personnel.

We recommend giving the product a thorough inspection when changing
location or patient.

Replacement components and accessories 

Worldwide ex US

US only

White Seat Covers

09 132 4557

Blue Seat Covers

09 132 4565

08 156 0325

Blue Antimicrobial Covers

09 143 4737

Standard Hand Controller 

09 132 5018

Premium/XL Handcontroller

09 143 4695

08 156 0358

Standard Battery Recharger

UK/Ireland

09 132 5034

Europe

09 132 5026

Australia/New Zealand

09 132 5042

Premium/XL Fast Battery Charger

UK/Ireland

09 143 4703

Europe

09 143 4711

Australia/New Zealand

09 143 4729

United States

08 156 0341

Replacement sucker feet (Pack of 4)

09 132 8350

Spare Actuator

09 132 8376

Spare XL Actuator

09 143 7516

Swivel Seat 

09 132 4532

08 156 0333

Swivel Transfer Seat 

09 132 4540

08 156 0366

Recycling

Bathlift, hand controllers and recharger must be disposed of correctly, if in doubt
contact your local refuse site. 

The bath seat is made from plastic; Hand controller contains electronic
components and Lithium Ion batteries; Actuator and recharger contain electronic
components; please separate and dispose of correctly.

Troubleshooting Guide

Specifications for baths / bath lifts (See 

Fig. 4 & 5

)

Descriptions

Dimensions

A   Bath underside length

>800mm

B   Bath underside width (distance between suction cups)

280 – 610mm

C  Bath width at 8 cm height (base plate)

> 420mm

D  Height of bath (With height adapter max. 2 cm (accessories))

<455mm (475mm)

E  Upper bath width (external)

<710mm

F  Upper bath width (internal)

<630mm

Problem

Possible causes

Remedy

Lift does not move

Connector in hand controller

Check connector (see Fig 8)

Battery is flat

See charging battery instructions

Actuator failure

Repair by authorized service centre

Faulty spiral cable

Replace hand controller

Lift only travels upwards

Battery is flat

See charging battery instructions

Red light on hand 

controller lights up

Battery is flat

See charging battery instructions

Battery will not charge

Faulty electronics

Check / Repair by authorized service centre

Faulty battery

Check / Repair by authorized service centre

Faulty power supply

Check / Repair by authorized service centre