6
Care and cleaning
Seat unit and back rest
•
After bathing rinse the bath lift (without covers) with warm water. If the bath lift
is heavily soiled, wash it down with commercially available cleansing
detergents and brushes.
•
Remove the seat and backrest covers.
•
Dry the product with a cloth.
•
Use a mild disinfectant if required.
•
Never clean the drive or hand controller with a high pressure washer.
•
Whilst cleaning the seat unit be careful of the mobile lifting scissors.
•
Do not use strong solutions of disinfectant.
•
Ensure that the bath lift is completely dry before storing it.
•
We recommend cleaning the Bathmaster Deltis thoroughly when changing
location or patient.
Hand controller and battery charger
•
Please read the hand controller instructions.
•
Prior to cleaning, the charger must be removed from the power supply plug.
•
Never dip the charger into water, spray wash or wash with water.
•
Clean the charger with a cloth.
•
Clean the hand controller with a damp cloth and then dry off.
•
Ensure that the bath lift is completely dry before storing it.
Seat and backrest covers
•
Once removed from the main unit, the seat and backrest covers can be
washed at 40°C (max) in the washing machine.
•
Do not dry the covers in a tumble dryer.
Technical specifications (See Fig. 2)
Total weight without hand control:
Deltis
|
Deltis XL ..................10.5kg
|
11.1kg
Weight of seat unit:
Deltis
|
Deltis XL...............................................6.8kg
|
7.4kg
Weight of back rest
.........................................................................................3.7kg
Weight of hand controller:
Deltis
|
Deltis Premium & Deltis XL ..........0.3
|
0.4kg
Seat depth (A):
............................................................................................500mm
Max seat depth (B):
.....................................................................................455mm
Min seat height:
.............................................................................................65mm
Total depth (back rest upright) (C):
............................................................690mm
Total depth (back rest lowered) (C):
..........................................................900mm
Total height (D):
.........................................................................................1105mm
Height of back rest (E):
...............................................................................680mm
Width of back rest (F):
................................................................................350mm
Width of seat with side flaps (G):
...............................................................710mm
Width of seat without side flaps:
................................................................375mm
Footprint of the base plate:
..............................................................530 x 220mm
Backrest reclines to:
...........................................................................................40°
Load capacity:
Deltis
|
Deltis XL ....................................................140kg
|
170kg
Actuator degree of protection against water ingress:
...................................IP66
Hand controller degree of protection against water ingress:
.......................IP67
Power on time:
.................................................................10%/2 min - 90%/18 min
Supply voltage:
Deltis
|
Deltis Premium, Deltis XL & Deltis US ......11.1V
|
14.8V
Operating voltage:
Deltis
|
Deltis Premium, Deltis XL & Deltis US...........5A
|
5A
Rechargeable Battery:
....................................................................................Li-ion
Charge time when low voltage indicated:
Deltis
|
Deltis Premium,
Deltis XL & Deltis US .............................................................................5h
|
1h
Decibel Level:
.................................................................From/To 49db(A)/47db(A)
Charger:
Deltis
|
Deltis Premium & Deltis XL
|
Deltis US
Input:
...............................................................................................230V AC/50Hz
..................................................................................
230 - 240V AC/50Hz
...............................................................................................
120V AC/60Hz
Output:
..................................................................................15V DC/500mA/7.5VA
................................................................................
20V DC/120mA/10VA
....................................................................................
20V DC/500mA/10VA
Warranty
Patterson Medical Ltd warrants that the Bathmaster Deltis Bath Lift is free from
defects in material and workmanship. This warranty shall remain in effect for
three years from the date of original consumer purchase.
In the event that the unit needs inspection or repair, contact Patterson Medical
Ltd, the distributor or an approved agent.
This warranty does not cover the following:
•
Replacement parts or labour furnished by anyone other than Patterson
Medical Ltd, the distributor or an approved agent.
•
Any failure of the unit during the warranty period if the failure is not caused by
defect in material or workmanship or if the failure is caused by unreasonable use.
Patterson Medical Ltd is not liable for incidental or consequential injury or
damage to property.
Should it be necessary to return the Bathmaster Deltis, it is essential it is
adequately packed to protect it in transit, preferably in its own purpose made
original box, otherwise Patterson Medical Ltd cannot accept any responsibility
for transit damage to the unit.
Any attempt at repair or maintenance by persons other than Patterson Medical
Ltd trained or approved personnel will void this warranty.
Maintaining the Unit
The rechargeable battery in the hand control is not interchangeable and may
only be opened by the manufacturer.
The Bathmaster Deltis should be inspected before each use to ensure it is free
from damage and excess wear. The Bathmaster Deltis is built to give long,
reliable use, without need for servicing. Whilst being robust, the lift should be
treated with care and in particular should not be dropped. If damage or
deterioration is suspected, please contact Patterson Medical Ltd or the
distributor, for repair by an approved agent.
For professionals issuing the Bathmaster Deltis, Patterson Medical Ltd
recommend that it receives bi-annual inspections in line with requirements of the
Lifting Operations and Lifting Equipment Regulations 1998 (LOLER). Inspections
should only be carried out by trained and experienced personnel.
We recommend giving the product a thorough inspection when changing
location or patient.
Replacement components and accessories
Worldwide ex US
US only
White Seat Covers
09 132 4557
Blue Seat Covers
09 132 4565
08 156 0325
Blue Antimicrobial Covers
09 143 4737
Standard Hand Controller
09 132 5018
Premium/XL Handcontroller
09 143 4695
08 156 0358
Standard Battery Recharger
UK/Ireland
09 132 5034
Europe
09 132 5026
Australia/New Zealand
09 132 5042
Premium/XL Fast Battery Charger
UK/Ireland
09 143 4703
Europe
09 143 4711
Australia/New Zealand
09 143 4729
United States
08 156 0341
Replacement sucker feet (Pack of 4)
09 132 8350
Spare Actuator
09 132 8376
Spare XL Actuator
09 143 7516
Swivel Seat
09 132 4532
08 156 0333
Swivel Transfer Seat
09 132 4540
08 156 0366
Recycling
Bathlift, hand controllers and recharger must be disposed of correctly, if in doubt
contact your local refuse site.
The bath seat is made from plastic; Hand controller contains electronic
components and Lithium Ion batteries; Actuator and recharger contain electronic
components; please separate and dispose of correctly.
Troubleshooting Guide
Specifications for baths / bath lifts (See
Fig. 4 & 5
)
Descriptions
Dimensions
A Bath underside length
>800mm
B Bath underside width (distance between suction cups)
280 – 610mm
C Bath width at 8 cm height (base plate)
> 420mm
D Height of bath (With height adapter max. 2 cm (accessories))
<455mm (475mm)
E Upper bath width (external)
<710mm
F Upper bath width (internal)
<630mm
Problem
Possible causes
Remedy
Lift does not move
Connector in hand controller
Check connector (see Fig 8)
Battery is flat
See charging battery instructions
Actuator failure
Repair by authorized service centre
Faulty spiral cable
Replace hand controller
Lift only travels upwards
Battery is flat
See charging battery instructions
Red light on hand
controller lights up
Battery is flat
See charging battery instructions
Battery will not charge
Faulty electronics
Check / Repair by authorized service centre
Faulty battery
Check / Repair by authorized service centre
Faulty power supply
Check / Repair by authorized service centre