Page 27
CONFIGURATION PARAMETERS
All the parameters that can be set are described in the following pages. At the end of each parameter
description, where present, the fieldbus address corresponding to the parameter is shown. If the pa-
rameter is the selectable type, the value to be entered in the register for the desired selection is shown
between “[ ]”.
CALIBRATION MENU
Access to this menu is only allowed when operating in FREE mode or when operating in METRIC with
access using the password of authorised personnel.
rESoLU
DIVISION VALUE [1101÷1102]
Value of a single division, expressed in kg. The ratio between the capacity of the system and the
division valuerepresents the resolution of the system (number of divisions).
Following the change of the division value, if the maximum capacity is not changed, calibration of
the weight is corrected automatically.
Values that can be selected:
0.0001 - 0.0002 - 0.0005
0.001 - 0.002 - 0.005
0.01 - 0.02 - 0.05
0.1 - 0.2 - 0.5
1 -2 - 5
10 - 20 - 50
Default: 1
L.C. CAP.
CAPACITY OF THE LOAD CELLS [1103-1104]
It defines the value corresponding to the sum of the rated capacity values of the load cells expressed
in kg. In the case of systems with only one load cell and “N” fixed supports, enter the capacity value
of the cell for the total number of supports. This figure represents the full scale value of the weighing
system. Following the change of the parameter value, the theoretical weight value is recalculated.
Values: from 1 to 999999
Default: 0
L.C. SEn..
SENSITIVITY OF THE LOAD CELLS [1105]
Set the value corresponding to the average sensitivity of the loadcells, in mV/V. The instrument accepts
values between 0.0 and 7.6 mV/V. If no value if programmed, 2mV/V is set by default.
Following the change of the sensitivity value, the theoretical weight value is recalculated.
Values: from 0.0000 to 4.0000 mV/V
Default: 2.0000