Page 54
UPLOAD/DOWNLOAD FUNCTION
The TESTER 1008 must be connected to the serial COM1 (RS232) of the instrument.
This feature allows you to download or upload the setup configuration and calibration data stored in
the instrument.
• Download function:
The instrument setup parameters are stored in a file.
• Upload function:
the instrument is configured with the setup parameters read from a file.
To use these functions it is necessary to activate the procedure ( “receive” files or “transmit file”) in
1008 TESTER instrument.
ACCESS VIEWING
This menu only appears in case of METRIC functioning.
SUB MENU
MESSAGGE
NAME
DESCRIZIPTION
TYPE
ACCEss
ACC-01
Access 01
Procedure for wiewing last access of
authorized personnel.
Spc
ACC-02
Access 02
Procedure for wiewing last but one access of
authorized personnel.
Spc
ACC-03
Access 03
Procedure for wiewing last but two access of
authorized personnel.
Spc
ACC-04
Access 04
Procedure for wiewing last but three access
of authorized personnel.
Spc
ACC-05
Access 05 Procedure for wiewing last but four access of
authorized personnel.
Spc
PROCEDURE FOR ACCES VIEWING
Press PRG to enter: you will see the operator code used to access the programming of the instrument
and the sequential number of accesses as indicated below:
1d.0000
The display shows the ID code used for access (the password table). In the case where access is made
through calibration jumper, it displays the identification code 0000.
Press the PRG key to continue with the visualization of the progressive access number.
Pr.0000
The display shows the sequence number of access (this value is incremented at each access and is
never reset). Press the PRG key to exit the access viewing procedure.