Drawing Number: 21075 
Revision: C 
ECN Number: 47023 

 

OPERATION MANUAL FOR 

QUARTZ PRESSURE SENSORS 

SERIES 102 
SERIES 113 

 
1.0 

INTRODUCTION 

 

This  series  of  miniature  dynamic  pressure  sensors  is 
specifically  designed  for  shock  tube  and  blast  wave 
measurements  and  for  other  applications  requiring 
very high frequency, near non-resonant response. 
 
The  term  used  to  describe  the  transient  response  of 
this  model  series  is  “Frequency  Tailoring”  and  it 
encompasses several mechanical and electrical design 
features 

coupled 

with 

stringent 

in-process 

fabrication/test  procedures  with  heavy  emphasis  on 
the shock tube as a tool. 
 

2.0 

DESCRIPTION 

 

Although  this  series  consists  of  sensors  with  three 
basic  mechanical  configurations  and  six  different 
sensitivities,  each  model  is  basically  similar  in 
internal design. 
 

 

Typical ICP

®

 Probe Style Sensor 

 
Each  utilizes  the  acceleration-compensated  Series 
113 quartz piezoelectric element coupled to a source 
follower  type  miniature  electronics.    (See  “General 
Guide  to  ICP

®

  Instrumentation,”  G-0001B,  for  a 

detailed description of the ICP

®

 concept.) 

 
 

 
 
The  figure  above  shows the components of the basic 
ICP

®

 probe, i.e. the piezoelectric element and the  

ICP

®

  source  follower  amplifier.    These  components 

are joined together as an inseparable sealed assembly 
at  the  factory.    Disassembly  should  not be attempted 
in the field. 
 
Series  113B2x  are  in  a  probe  configuration  and  are 
installed  with  a  hollow  clamp  nut  with  5/16-24 
external  threads.    The  housing  of  these  models  is  at 
electrical ground potential. 
 
Series  113B3x  are  similar  to  the  B2x  Series  with  an 
additional  feature;  all  Invar  construction.    The  all 
Invar  sensors  are  designed  to  have  minimal 
susceptibility  to  thermal  transient  events  and  are 
specifically  suited  for  high-temperature  shock  and 
blast measurements. 
 

 

Series 113:  Probe Style Sensor 

 
Series 102 consist of the basic 113 Series probe, as in

 

the  above  mentioned  series,  mounted  in  a  3/8-24 
threaded mounting adaptor.  The probe is installed at 
the  factory  in  an  "off  ground"  configuration,  i.e.  the 
probe  body  is  insulated  from  the  external  mounting 
adaptor body. 
 
The  Model  102A12  utilizes  the  same  inner  probe 
design  as  the  above  two  designs  but  in  a  3/8-24 

+

_

Intergrated
Circuit
Amplifier

Seal Ring

Potting

Connector

Preload Sleeve

Electrodes

Diaphragm

Housing 0.218 Dia

Quartz Plates

Acceleration
Compensation
Mass and Plate

Mounting
Clamp Nut

M

 

 

Summary of Contents for 113B26

Page 1: ... Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Page 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ... shows the components of the basic ICP probe i e the piezoelectric element and the ICP source follower amplifier These components are joined together as an inseparable sealed assembly at the factory Disassembly should not be attempted in the field Series 113B2x are in a probe configuration and are installed with a hollow clamp nut with 5 16 24 external threads The housing of these models is at ele...

Page 7: ...ial in many cases One or more layers may be used across the end of diaphragm and adaptor A silicone rubber coating approximately 010 thick has also been proven effective in many applications General Electric RTV type 106 is recommended Apply the rubber coating to the surface of the diaphragm and allow it to cure in accordance with the manufacturer s instructions If you have ordered the CA option a...

Page 8: ...s are also features of these sensors As mentioned previously the diaphragm must be flush mounted to fully realize the high frequency response capabilities of this series 8 0 CALIBRATION Piezoelectric sensors are dynamic devices but static calibration means can be employed if discharge time constants are sufficiently long Generally static methods are not employed below several hundred seconds time ...

Page 9: ......

Page 10: ......

Reviews: