background image

 

PCB Load & Torque   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcb.com

 

 

 

CANISTER LOAD CELL OPERATION MANUAL 

 

4

 

6.1.1 Measured Output 

 

The  applied  load  starting  at  zero  is  measured  in  five 
increments  to  full  scale.  Output  (mV/V)  is  measured  at  each 
increment.  The  straight-line  from  zero  to  the  full  scale 
measurement  is  compared  to  the  measured  readings  at  each 
increment  to  calculate  the  error  at  each  load  increment.    The 
deviations  (%  Full  Scale)  corresponding  to  non-linearity  at 
each measurement increment are then calculated. 

6.1.2 Hysteresis 

 

The  difference  between  the  ascending  and  descending 
measured readings at 40% of full scale is used to calculate the 
hysteresis value. 

6.1.3 Best Fit Output 

 

The  best  fit  calibration  second-order  equation  has  been 
calculated  from  the  calibration  data  by  the  method  of  least 
squares.  Deviation  between  measured  output  and  best-fit 
output  is  calculated  and  displayed  in  the  column  next  to  the 
best-fit  output  for  each  measurement  increment.  The 
deviations  (%  Full  Scale)  of  measured  outputs  from  the 
calculated best fit are tabulated for each measured reading. 

6.1.4 Strain Gage Measurements 

 
 
 
 
 
 
 
 
 
 

6.1.5 Shunt Calibration Standard Resistor 

 

All  canister  load  cell  calibrations  use  a  120K  Ohm  (0.1%) 
precision  resistor  shunt  calibration  value  that  is  supplied  into 
the calibration report. 

6.1.6 Static Error Band (SEB) 

 

The  static  error  band  (SEB)  is  determined  by  the  maximum 
deviations of  the ascending and descending calibration points 
from  the  best  fit  straight  line  through  zero  output.  The  SEB 
includes the effects of nonlinearity, hysteresis, and non-return 
to minimum load. 
 
 
 
 

7.0 SHUNT CALIBRATION DESCRIPTION 

 
Shunt  calibration  is  used  to  simulate  a  known  tension  or 
compression  load  on  a  load  cell.  The  calibration  certificate 
will indicate which leg of the bridge to apply the shunt resistor 
to for both tension and compression load simulation. Typically 
tension is simulated by inserting the shunt resistor between the 
+P and +S connector leads. Compression loading is simulated 
by  inserting  the  shunt  resistor  between  the  +S  and  –P 
connector leads. 
 

7.1 Resistor Value 

Canister load cells have a nominal 2.0 mV/V full scale output. 
For a 700 ohm strain gage bridge the precision shunt resistor, 
120,000  ohms 

±

  0.1%,  simulates  an  output  of  approximately 

73% of the  full scale output  for the load cell. The calibration 
values for each bridge are found on the calibration certificates 
supplied with each load cell. 
 

7.2 Shunt Calibration Process 

To perform the shunt calibration, use the following procedure: 
 

1.

 

Stabilize  all  forces  on  the  load  cell.    If  possible, 

remove all loads. 

 
2.

 

Power up the host signal conditioner and connect it to 

the load cell via appropriate cable, and allow for a 30 
minute warm up. 

 

3.

 

Set the load indicator display to read exactly 00.000. 

 
4.

 

Connect  the  shunt  resistor  to  the  terminals  specified 

in  the  calibration  certificate,  and  adjust  the  span  or 
gain until the display reads the force value stated on 
the certificate. 

 
5.

 

Repeat  steps  1-3  to  verify  that  a  valid  calibration 

setting has been obtained. 

 
6.

 

If  possible,  apply  a  known  load  to  the  measurement 

system to further verify that the calibration has been 
accurately set up. 

7.3 Estimating Shunt Resistor for a Given Load 

The  following  formula  can  be  used  to  estimate  the 
approximate  value  of  shunt  resistor  required  to  simulate  a 
mechanical load. 
 
 

R

cal

 = (25 * R

b

) / (Output

FS

 * L

cal

)   

 
Where: 

R

cal

 = Shunt Resistor (K ohms) 

 

R

b

 = Bridge Resistance (ohms) 

 

Output

FS

 = Full Scale output of the load cell (mV/V) 

 

L

cal

 = Load to be simulated, % of Load Cell Capacity 

Bridge Resistance: 

700 Ohm Nominal 

Excitation: 

+P(A) to –P(D) Ohms 

Signals: 

+S(B) to –S(C) Ohms 

Leakage to Ground: 

> 5k GOhm 

Bridge Unbalance: 

±

1.0% Full Scale 

Output: 

2.0 mV/V Nominal 

Maximum Voltage: 

20 VDC 

 

Table 5

 

– Strain Gage Measurements 

Summary of Contents for 1102-01A

Page 1: ...nstallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbloadtorque com Web www pcbLoadTorque com ...

Page 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Page 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...ing Considerations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 4 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Standard Resistor 4 6 1 6 Static Error Band SEB 4 7 0 SHUNT CALIBRATION DESCRIPTION 4 7 1 Resistor Value 4 7 2 Shunt Calibration Process 4 7 3 Estimating Shunt Resistor for a Given Load 4 8 0 MAI...

Page 7: ...nts 4 0 MECHANICAL INSTALLATION Loads applied to canister load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing fixtures be sure to thread the fixture to the load cell and not the load cell to the fixture If alignment fixtures are not used ensure loading surfaces are flat and parallel wit...

Page 8: ...instrument or control system is grounded through its power cord Ground loops and measuring system wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potentia...

Page 9: ...leg of the bridge to apply the shunt resistor to for both tension and compression load simulation Typically tension is simulated by inserting the shunt resistor between the P and S connector leads Compression loading is simulated by inserting the shunt resistor between the S and P connector leads 7 1 Resistor Value Canister load cells have a nominal 2 0 mV V full scale output For a 700 ohm strain ...

Page 10: ...e signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 10 0 CALIBRATION REPAIR SERVICES PCB Load Torque offers calibration and repair services The PCB Calibration Laboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point des...

Page 11: ...x 38 1 mm 7 Weight 0 67 lb 304 g Mounting Thread 1 4 28 Female No Metric Equivalent Housing Material Painted Aluminum Painted Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 003 in 0 08 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve ...

Page 12: ...ULL 1102 SERIES PT CONN CANNISTER LOAD CELL PTE 9 14 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 51277 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque Inc REVISIONS REV DESCRIPTION ECO A ADDED NOTE 1 WIRING CODE 38165 PT02E 10 6P RECEPTACLE T...

Reviews: