background image

 

PCB Load & Torque, Inc.   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcbloadtorque.com

 

 

 

S-TYPE LOAD CELL OPERATION MANUAL 

 

3

 

5.1 Electrical Drawing / Western Regional Std. 

All  load  cells  are  wired  following  the  Western  Region 
Standard.    All  models  utilize  strain  gages  configured  into  a 
Wheatstone  Bridge  Circuit  to  produce  the  primary  sensing 
element.  The  four-arm  Wheatstone  bridge  configuration  is 
shown below in Figure 3. 

 
 
 
 
 
 
 
 

The  gages  are  bonded  to  the  load  cell's  structure. 
Typically,  a  regulated  DC  or  AC  excitation  is  applied 
between A and D of the bridge. When a force is applied to 
the  load cell,  the Wheatstone  bridge  becomes  unbalanced, 
causing  an  output  voltage  between  B  and  C,  which  is 
proportional to the applied load.  This configuration allows 
for  temperature  and  pressure  compensation,  as  well  as 
cancellation  of  signals  caused  by  forces  not  directly 
applied  to  the  axis  of  the  applied  load.  Output  is  typically 
expressed in units of mV/V of excitation. 
 

5.2 Output Polarity 

The following figure describes the output polarity. 

 

 

 

 
 
 
 
 
 

 

 
 
 

5.3 Cable & Grounding Considerations 

Proper  grounding  and  shielding  is  required  to  prevent 
electrical  noise  in  strain  gage  load  cell  measuring  systems.  
The cable must be shielded twisted pairs with a drain wire. 
 
Cable shields must be grounded only at one end, for example, 
on  the  instrument  or  control  system  ground.    The  load  cell 
case is grounded by mechanical attachment to the structure to 
which it is mounted. 
 

The  instrument  or  control  system  is  grounded  through  its 
power cord.  Ground loops and measuring system wiring may 
result in unstable or noisy signals. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
A  simple  test  with  a  voltmeter  connected  between  the  power 
cord  ground  and  the  structure  on  which  the  load  cell  is 
mounted  can  confirm  that  the  structure  has  been  properly 
grounded.  If the power cord ground and structure ground are 
not  at  the  same  potential,  it  may  be  necessary  to  provide  a 
secure  structure  ground,  perhaps  by  driving  a  copper  rod  and 
attaching a ground strap. 
 
 

6.0 CALIBRATION 

 

Every S-type load cell manufactured by PCB Load & Torque, 
Inc. has been  fully calibrated  per ISO/IEC 17025 procedures, 
and  meets  all  published  specifications.    Each  load  cell  will 
come  with  a  calibration  certificate  designated  with  matching 
model  and  serial  numbers.  PCB  Load  &  Torque  also  offers 
calibration services on an on-going basis. 

 
 

6.1 Calibration Certificate Description 

Calibration  reports  supplied  with  S-type  load  cells  contain 
valuable  information  to  assist  the  customer  in  use  of  the 
equipment.  A  separate  calibration  report  is  provided  for 
tension  and  compression  calibrations.  Calibration  procedures, 
equipment, and reports comply with ISO/IEC 17025. 

6.1.1 Measured Output 

 

The  applied  load  starting  at  zero  is  measured  in  five 
increments  to  full  scale.  Output  (mV/V)  is  measured  at  each 
increment.  The  straight-line  from  zero  to  the  full  scale 
measurement  is  compared  to  the  measured  readings  at  each 
increment  to  calculate  the  error  at  each  load  increment.    The 
deviations  (%  Full  Scale)  corresponding  to  non-linearity  at 
each measurement increment are then calculated. 
 

Figure 5

 

- Grounding 

 

Test Rig Ground 

Power Supply 

Instrument Ground 

Cable Shielding 

Load Cell 

 

- Signal White 

- Excitation Black 

+ Excitation Red 

+ Signal Green 

Figure 3 - Western Regional Strain Gage Committee Wiring Code 

Figure 4

 

- Output Polarity 

 

Tensile Load 

= Positive 

Tensile Load 

= Positive 

Summary of Contents for PCB Load & Torque 1630-03C

Page 1: ...stallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbl...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...erations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 3 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Stand...

Page 7: ...STALLATION Loads applied to S type load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing...

Page 8: ...or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded I...

Page 9: ...ision shunt resistor is 60 000 ohms 0 1 simulates an output of approximately 73 of the full scale output for the load cell The calibration values for each bridge are found on the calibration certifica...

Page 10: ...aboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point descending Additional data points are avail...

Page 11: ...104 g Mounting Thread 1 4 28 thread No Metric Equivalent Housing Material Aluminum Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 0025 in 0 05 mm Electrical Conn...

Page 12: ...6 23 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 46137 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed...

Reviews: