13
n
C160 : 1000 A calibre (2000 A max. peak)
Intensity in A~
1...50 A
50...200 A
200...1000 A
1000...1200 A
Intrinsic error
(1)
≤
1% + 1 mV
Dephasing
(2)
≤
3°
≤
2°
≤
1°
C160 : 100 A calibre (300 A max. peak)
Intensity in A~
0.1...5 A
5...20 A
20...100 A
100...120 A
Intrinsic error
(1)
≤
2% + 5 mV
Dephasing
(2)
≤
15°
≤
10°
≤
5°
C160 : 10 A calibre (30 A max. peak)
Intensity in A~
0.1...0.5 A
0.5...2 A
2...10 A
10...12 A
Intrinsic error
(1)
≤
3% + 10 mV
Dephasing
(2)
(2)
≤
15°
≤
15°
(1) As % of output signal
(2) Unspecified
n
C173 : 1000 A calibre
Intensity in A~
1...10 A
10...100 A
100...1000 A
1000...1200 A
Intrinsic error
(1)
≤
1% + 0.2 mV
≤
0,5% + 0.2 mV
≤
0.2%
≤
0.2%
Dephasing
(2)
≤
2°
≤
1°
≤
1°
C173 : 100 A calibre
Intensity in A~
0.1...1 A
1...10 A
10...100 A
100...120 A
Intrinsic error
(1)
≤
1% + 0.2 mV
≤
0.5% + 0.2 mV
≤
0.3%
≤
0.2%
Dephasing
(2)
≤
2°
≤
1°
≤
1°
C173 : 10 A calibre
Intensity in A~
0.01...0.1 A
0.1...1 A
1...10 A
10...12 A
Intrinsic error
(1)
≤
1% + 0.2 mV
≤
0.5% + 0.2 mV
≤
0.5%
≤
0.5%
Dephasing
(2)
≤
5°
≤
2°
≤
2°
C173 : 1 A calibre
Intensity in A~
0.001...0.01 A
0.01...0.1 A
0.1...1 A
1...1.2 A
Intrinsic error
(1)
≤
3% + 1 mV
≤
3% + 1 mV
≤
0.7% + 1 mV
≤
0.7% + 1 mV
Dephasing
(2)
(2)
≤
10°
≤
10°
(1) As % of output signal
(2) Unspecified
4.4
Conditions of use
The "C100" clamps must be used in the following conditions, in order to comply with user safety and
metrological performance requirements.