If an over-range or under-range condition has occurred at
least  once  during  measurement  the  “  OV  ”  or  “  UN  ”
indication  is  shown  on  the  display,  to  indicate  that  over-
range or under-range data were included in the sound level
measurement values for processing.
During  measurement,  if  change  the  frequency  or  time
weighting  and  sound  level  ranges,  the  stored  maximum
and  minimum  sound  level  will  be  clear.    The  maximum
recorded elapsed time is 100 hours.

10-2 Store operations

1. To memorize the reading

  Press  “MEM”  key  each  time  to  store  one  set

measurement  data  (L,  Lmax,  Lmin  and  each  elapsed
time) in memory, and LCD will show “ M ” and memory
location numbers (1 to 99).

 When the data number 99 is reached, the “FULL”

indication is shown on the display, does not change
further and does not return to 1.

2. To recall and read the reading

 Press "READ" key to recall the reading memory data

mode. LCD will show "    READ    " and memory
location numbers.

  Press  “

”  or  “

 

 

  key  to  scroll  through  the  logged

readings.

  Press  “MAX”  key  to  cycle  switch  the  stored  currently

sound  level  with  elapsed  measurement  time 

  the

maximum sound level (“max” is shown) with measured
elapsed  time 

  the  minimum  sound  level  (  “min”  is

shown) with measured elapsed time.

 Press "READ" key again to exit READ mode.

-12-

3. To clean the memory

 Press “ 

 ” key to turn-off the meter.

 Press and hold down “MEM” key then press “ 

 ” key

to turn on the meter, LCD will show “CLr” and all stored
data are clear.

OUTPUT CONNECTORS

3.5

φ

 Plug Wiring Connection

Dc output

Ac output

Common output

AC Output :

An AC signal corresponding to the frequency-weighted signal
is output.

Output voltage : 1Vrms

±

100mVrms (scal upper limit)

Output impedance : approx. 5K

Ω

Load impedance :

1M

Ω

The output voltage when the meter is in calibration mode
(-6dB from scale upper limit, 1000Hz sine wave) is 0.5Vrms.

DC Output :

A level-converted DC signal generated by rms detection and

logarithmic compression is output.    The signal reflects the
frequency and time weighting settings of the meter.

Output voltage : 10mV

±

0.1mV/dB

Output impedance : approx. 5K

Ω

Load impedance :

1M

Ω

The output voltage when the meter is in 94dB reading is
0.94VDC.

-13-

Summary of Contents for PCE-318

Page 1: ...1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info industrial needs com ...

Page 2: ...not covered in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instruments Safety symbols Comply with EMC U S Pat No Des 483 680 Ⅱ Ⅱ Ⅱ Ⅱ PRECAUTIONS Do not try to remove the microphone grid you can easily damage the microphone in this way Protect the meter from impact Do not drop it Transport it in the...

Page 3: ...in with elapsed time Both AC and DC signal output is available from a single standard 3 5mm coaxial socket suitable for a frequency analyzer level recorder FFT analyzer graphic recorder etc 2 Ⅳ Ⅳ Ⅳ Ⅳ PARAMETERS The following parameters are used on the meter LA A frequency weighting sound pressure level LC C frequency weighting sound pressure level FAST Fast time weighting SLOW Slow time weighting ...

Page 4: ...resolution Bar graph indication of current sound level 1dB resolution Elapsed time display 100 hours maximum Memory and read address display 99 data sets maximum 4 Warning indications Over range indication OV displayed at upper limit of the range Un displayed at lower limit of the range Outputs AC output using selected frequency weighting Output voltage 1Vrms at full scale of the range Output impe...

Page 5: ... key Press to read the stored data sets in memory press again to exit read mode 9 keys Level range keys select the level range for the measurement The following six settings are available 20 to 80 30 to 90 40 to 100 50 to 110 60 to 120 70 to 130 When read mode the keys select the memory address to be displayed 10 DC AC output AC signal with frequency weighting DC signal corresponding to sound leve...

Page 6: ...y and before connecting the AC adaptor be sure to turn off the meter 1 Battery Loading Before replace the batteries must press key turn off the meter Remove the cover of the battery compartment Insert the new 9V battery Replace the battery cover Press key turn on the meter then press key turn off the meter 2 Battery capacity indicator When use battery operating the meter periodically check this in...

Page 7: ... fast time weighting and 60 to 120dB sound level range 2 Press A or C keys to select desired frequency weighting For normal sound level measurements select A setting 3 Press FAST or SLOW keys to select desired time weighting dynamic characteristics Normally the FAST setting should be used 4 When performing measurements according to IEC or other standards the frequency and time weighting setting re...

Page 8: ...h the logged readings Press MAX key to cycle switch the stored currently sound level with elapsed measurement time the maximum sound level max is shown with measured elapsed time the minimum sound level min is shown with measured elapsed time Press READ key again to exit READ mode 12 3 To clean the memory Press key to turn off the meter Press and hold down MEM key then press key to turn on the met...