5.3.4 Environment Set
In this menu you can set whether or not the
temperature and humidity are displayed. Select
Enable to display the air temperature and humidity
or Disable to switch this function off.
5.3.5 Sample Cycle
In this menu you can set the number of
measurements.
5.3.6 Sample Mode
In this menu you can adjust the display mode of the
particle counter.
Cumulative -
Differential -
Concentration -
5.3.6.1 Sample modes explanation
-
Cumulative and differential:
The recorded numbers of particles are added up per channel (cumulative); starting with the
largest channel
In differential mode there are the absolute particle numbers per channel
Example:
Channel [µm]
Cumulative particle number
[PC]
Differential particle number [PD]
Calculation
0.3
1630
1000
= [PC,630]+[PD,1000]=[PK,1630]
0.5
630
350
= [PC,280]+[PD,350]=[PK,630]
1.0
280
200
= [PC,80]+[PD,200]=[PK,280]
2.5
80
50
= [PC,30]+[PD,50]=[PK,80]
5.0
30
20
= [PC,10]+[PD,20]=[PK,30]
10.0
10
10
---
-
Concentration:
Here, you can find the average numbers of the respective particles per channel; also called
particle density.