background image

 

©  PCE Instruments 

 

 
 
 
 
 

 

 

PCE-CT 5000H Coating Thickness Gauge 

User Manual 

User manuals in various languages (français, 

italiano, español, português, nederlands, türk, polski, 

русский, 

中文

) can be found by using our 

product search on: www.pce-instruments.com 

Last change: 6 September 2018 

v1.0 

E

nglis

h

 

Summary of Contents for 5000H

Page 1: ...ing Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last ch...

Page 2: ...5 2 Basic Measurement Step 5 5 3 User Interface 6 5 4 Measure Mode Only Support Single Mode 7 5 5 Group Mode 7 5 6 Statistics 7 5 7 Probe Mode 8 5 8 Storage 8 5 9 Recall and Delete Readings 8 5 10 Hi...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...designed for non destructive fast and precise coating thickness measurement The principal applications lie in the field of corrosion protection It is ideal for manufacturers and their customers for o...

Page 5: ...de 2 High and low limit alarm 3 Probe mode AUTO MAG EDDY 4 Measuring readings display 5 Statistics display MIN MAX AVG 6 Auto power off indication 7 USB connecting indication 8 Measuring unit 9 Low ba...

Page 6: ...surement no measurement data storage four data groups automatic measurement data storage of up to 2000 measured values Statistical functions Number of measurements average minimum maximum standard dev...

Page 7: ...tery 2 The gauge will work in factory default settings for the first time including single measure mode AUTO probe mode direct group mode DIR etc 3 When power on if in direct group mode DIR the readin...

Page 8: ...Exit b Switch on off the backlight in measure mode Up Key a Move up or roll up b Increasing c Delete the last readings or whole group Down Key a Move down or roll down b Decreasing c Refresh LCD on t...

Page 9: ...thod of switch group mode a Press Left Key once to enter into menu mode Root directory b Press Up Key or Down Key to activate Options item and press Left Key once to enter in c Press Up Key or Down Ke...

Page 10: ...l group mode GENn readings and statistics will be stored and not be lost when power off In direct group mode DIR readings and statistics will be lost when power off Each group has individual high low...

Page 11: ...en you switch the work group the applied alarm settings will also be switched automatically Set high low limit for the current work group as below a Press Left Key once to enter menu mode Root directo...

Page 12: ...ss Up Key or Down Key to select intended item Press Left Key to confirm it 5 13 Data Download User can use the USB cable to connect the gauge with PC and download stored readings If USB connected the...

Page 13: ...asuring accuracy 6 Factors Affecting Measurement Accuracy User needs to know the factors of affecting measurement accuracy before using the gauge The factors are listed below Factors Principle Magneti...

Page 14: ...recalibrate the gauge according to practical applications 7 Calibration User should read chapter 6 Factors Affecting Measurement Accuracy firstly and must recalibrate the gauge according to measured...

Page 15: ...individual zero calibration and multi point calibration b When calibration up to 4 points user must clear the finished points firstly to restart c Readings measured before will not be affected by new...

Page 16: ...roup and the other groups will not be affected 8 Troubleshooting If the instrument gives no response and can t be turned on you can remove the battery and a few minutes later reinstall the battery and...

Page 17: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 18: ...975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www...

Reviews: