Turbidity Calibration 

 

The PCE-TUM 50 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use. 

 

During the calibration process, do not shake or agitate the calibration standard violently to prevent air bubbles. 

 

In order to get accurate measuring results, we recommend calibrating the meter at least once a month.

 

The  meter  allows  turbidity  calibration  up  to  7  points  with  minimum  of  2  points,  the  default  calibration  points  include  the  0.02,  10.00,  200,  500,
1000, 1500 and 2000 NTU. You are able to select the calibration point during the calibration.

Calibrate the meter 

1.

Press the 

Cal

 key to enter the calibration mode, the meter prompts to

select   

the “Number of Calibration Points”.

2.

Press  the 

  or 

  key  to  select  an  applicable  option (2  to  7  points),

press the   

Enter

 key, the display shows 0.02 NTU.

 

3.

If necessary, press the 

 or 

 key to select desired calibration point

(e.g.,   10.00 NTU). 

4.

Insert  the  corresponding  standard  into  the  sample  chamber,  ensure

that the   

mark on the vial cap always aligns with the arrow on the meter. 

Close the light   

shield lid.

5.

Press  the 

Enter

  key,  the  meter  begins  the  calibration.  The 

Calibrating..icon    shows in the upper right corner of screen.

6.

Wait for the measured value to stabilize, the display automatically show 

the 

next calibration point (e,g., 200 NTU). 

7.

Repeat  steps  3  to  5  above  until  the  display  shows  “Calibration  is

Completed”. 

 

The meter returns to the measurement mode. Calibration

is completed.

 

During the calibration process, press the 

ESC

 key, the meter will exit the calibration and return to the measurement mode.

Summary of Contents for PCE-TUM 50

Page 1: ...PCE TUM 50 Turbidity Meter Instruction Manual...

Page 2: ...npacking ensure that the current work environment meet following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic an...

Page 3: ...l down through the menu item ENTER Confirms the calibration settings or displayed options MEAS Starts measurement Locks the measured value Connecting the Power Adapter Before plugging in the power ada...

Page 4: ...ale of the American Society of Brewing Chemists mg L Total Suspended Solids Unit Resolution 0 1 Set the resolution of the turbidity measurement 0 1 0 01 TSS Factor Set the conversion factor for measur...

Page 5: ...Solids mode is selected the measurement unit will switch to mg L 2 1 Press and hold the key for 3 seconds to enter the setup menu 2 2 Press the or key to select the Measurement Mode press the Enter k...

Page 6: ...n to the measurement mode Clear the stored data Please refer to page 13 Clearing the Memory Password The password protection is used to prevent the unauthorized calibration and settings If enabled the...

Page 7: ...default settings If enabled all of the calibration data and selected options parameters will be lost or reset the meter must be recalibrated 7 1 Press and hold the key for 3 seconds to enter the setu...

Page 8: ...rors The outside of the vial must be clean and dry before starting the test Wipe the vial with a lint free cloth to remove fingerprints or waterdrops If the vial has minor scratches or scuffs add a fe...

Page 9: ...select the Number of Calibration Points 2 Press the or key to select an applicable option 2 to 7 points press the Enter key the display shows 0 02 NTU 3 If necessary press the or key to select desire...

Page 10: ...ed for 1 minute so that bubbles can disappear 1 4 Wipe the vial with the lint free cloth to remove waterdrops and fingerprints Ensure that the outside of the vial is dry and clean 1 5 Place the vial i...

Page 11: ...pre weighed filter disk 3 6 Put the filter disk and watch glass in a 104 C drying oven for 1 hour 3 7 Remove the filter disk and watch glass and put in a desiccator Immediately cover the desiccator W...

Page 12: ...me as the turbidity Please refer to page 10 Turbidity Measurement Storing and Recalling Data The PCE TUM 50 turbidity meter is capable of storing and recalling up to 200 data sets Storing readings int...

Page 13: ...nd hold the key for 3 seconds to enter the setup menu 2 Press the or key to select the Delete Data press the Enter key to confirm 3 Press the or key to select the Delete all stored data 4 Press the En...

Page 14: ...ystem will automatically scan an available communication port and show the message box Found a port on your computer Click the OK button the application starts Click the Connect button the screen show...

Page 15: ...96 H mm Weight 1 2kg Addendum 1 Indexing and Matching Sample Vials The United States Environmental Protection Agency U S EPA recommends that vials used for turbidity calibration or sample measurement...

Page 16: ...l of hydrazine sulfate and 5 0ml of hexamethylenetetramine solutions in a 100ml volumetric flask and let stand 24 hours at 25 3 C 1000 NTU mix 25ml of 4000 NTU standard in a 100ml flask dilute to the...

Reviews: