9
H. Auto-Power-Off
The meter will turn itself off 30 minutes after power-on. To turn it on
again, user can either press the HOLD button or turn the power off and
turn it on again. To disable the Auto-power-off function, hold the HOLD
button while turning on power.
IV.Specifications
(23
℃±
5
℃
)
AC current:
Accuracy
Range
Resolution
50/60 Hz
Wide(400-1KHz)
40mA
10
μ
A
±
1.0%
±
0. 5mA
±
4.5%
±
0. 5mA
400mA
100
μ
A
±
3.0%
±
5.0mA
±
3.0%
±
5.0mA
4A
1mA
±
4.0%
±
0.1A
±
4.0%
±
0.1A
40A
10mA
±
4.0%
±
1.0A
±
4.0%
±
1.0A
80A
100mA
±
2.5%
±
1.0A
±
3.0%
±
1.5A
80-100A
1
100mA
±
5.0%
±
1.0A
±
5.0%
±
1.5A
1
Though the meter can display up to 400A,it is not calibrated beyond 100A
AC Voltage
( input impedance:10M
Ω
)
Overload
protection
Range
Resolution
50/60Hz
40-1KHz
400V
0.1V
±
1.5%
±
2dgts
±
2.0%
±
4dgts
AC 600V
Resistance (
Ω
) and continuity:
(open voltage 0.4V)
Range
Resolution
Accuracy
Beeping
OL Protection
40-400
Ω
0.1
Ω
±
1.0%
±
2dgts
<38.0
Ω
AC 600V
Conductor size:
30mm max.(approx.)
Battery Type:
One 9V NEDA 1604
Display:
3 3/4 LCD with 40seg. bargraph
Range Selection:
manual
Overload indication:
left most digit blinks
Power Consumption: 10mA(approx.)
Low battery indication:
Sampling Time:
2 times/sec.(display)
20 times/sec.(bargraph)