info@tursdaletechnicalservices.co.uk
5
1.2 Electrical Specifications
Accuracies are specified in the way:
±(…% of r…digits) at 23ºC±5ºC,below 80% RH.
OHMS
Range Resolution
Accuracy
Max. open
Circuit Voltage
Overload
Protection
40.00
Ω
0.01
Ω
+(1.2%+3)
5.8V
250Vrms
400.0
Ω
0.1
Ω
5.8V
Continuity Beeper
Range Resolution Operation
Resistance
Max.open Circuit
Voltage
Overload
Protection
•))) 0.01
Ω
Resistance
≤
35
Ω
5.8V
250Vrms
Short circuit current
≧
200mA
DC Voltage
Range Resolution
Accuracy
Input Impedance
Overload
Protection
1000V 1V
+(0.8%+3)
10M
Ω
1000Vrms
AC Voltage (40Hz~400Hz)
Range Resolution
Accuracy
Input Impedance
Overload
Protection
750V 1V
+(1.2%+10)
10M
Ω
750Vrms
Meg OHMS
Terminal
Voltage
Range Resolution
Accuracy
Test Current Short circuit
current
125V(0%~+1
0%)
0.125~4.000 M
Ω
0.001M
Ω
+(2%+10)
1mA
@load125k
Ω
≤
1mA
4.001~40.00 M
Ω
0.01M
Ω
+(2%+10)
40.01~400.0 M
Ω
0.1M
Ω
+(4%+5)
400.1~4000 M
Ω
1M
Ω
+(5%+5)
250V
(0%~+10%)
0.250~4.000 M
Ω
0.001M
Ω
+(2%+10)
1mA
@load250k
Ω
≤
1mA
4.001~40.00 M
Ω
0.01M
Ω
+(2%+10)
40.01~400.0 M
Ω
0.1M
Ω
+(3%+5)
400.1~4000 M
Ω
1M
Ω
+(4%+5)
500V(0%~+1
0%)
0.500~4.000 M
Ω
0.001M
Ω
+(2%+10)
1mA
@load500k
Ω
≤
1mA
4.001~40.00 M
Ω
0.01M
Ω
+(2%+10)
40.01~400.0 M
Ω
0.1M
Ω
+(2%+5)
400.1~4000 M
Ω
1M
Ω
+(4%+5)
1000V
(0%~+10%)
1.000~4.000 M
Ω
0.001M
Ω
+(3%+10)
1mA
@load1M
Ω
≤
1mA
4.001~40.00 M
Ω
0.01M
Ω
+(2%+10)
40.01~400.0 M
Ω
0.1M
Ω
+(2%+5)
400.1~4000 M
Ω
1M
Ω
+(4%+5)
All manuals and user guides at all-guides.com