126
Err.CAL
Calibration parameters lost – send the transducer
to a service, the message does not prevent measured
values from being displayed, message is displayed
in cycles.
Err Batt
Service
Real time clock battery low voltage – loss of real time clock
presets after a power loss, the transducer can operate,
consider sending the transducer to a service to replace
the battery; the message does not prevent measured
values from being displayed, message is displayed in cycles.
Changing date or hour settings switches of that message.
Err.PAR
Parameter error – restore default settings, do not ope
-
rate the transducer until default settings are restored,
the message does not prevent measured values from being
displayed, message is displayed in cycles.
Err Ind
Wrong configuration of individual characteristic parameters,
the transducer can operate – individual characteristic fun
-
ction does not work, the message does not prevent mea-
sured values from being displayed, message is displayed
in cycles.
Error
Par.File
Reading configuration from file stored on an external
SD/SDHC card or on the internal file system memory
unsuccessful – file is missing or corruped, the transducer
can be operated, the message does not prevent measured
values from being displayed, message is displayed in cyc-
les for about 20 seconds.