OPERATING PROCEDURE 
 
STANDARD MODE: 

 

1.

 

INSTRUMENT AND PROBE 

 

 

PLUG THE PROBE TO THE THICKNESS GAUGE CONNECOTR. PRESS

  KEY, HEAR BI-BI 

SOUND THEN LCD DISPLAY TURN ON AND DISPLAY THE LAST SET VERLOCITY. DISPLAY AS FOLLOWING 

 

 
 
THICKNESS GAUGE SYSTEM SETTING 

SYSTEM SETUP 

PRESS GAUGE

  KEY TURN ON THE GAUGE, LCD WILL DISPLAY 

 

PRESS MENU KEY 2 TIME. THEN PRESS ENTER KEY

, DISPLAY FOLLOWING 

 
System          Setup 

Function        Setup 
Outside          alarm 
[enter]    to    select 

↓to down     

 

PRESSS ENTER KEY 

AGAIN, DISPLAY FOLLOWING SCREEN 

 

Units   

 

METRICS    ( CAN PRESS 

SELECT INCH OR METRICS UNIT ) 

Resolution   

HIGH           ( HIGH=HIGH RESOLUTION DISPLAY 0.01 mm )     

Min Capture            ON          ( CAPTURE MIN MEASUREMENT THICKNESS ) 
2-Point CAL            OFF            ( OFF=SINGLE POINT AND ON=2-POINT CAL ) 
LANGUAGE            ENGLISH 

↓to down 
 

USE 

 

KEY MOVE HIGH-LIGHT CURSOR UP AND DOWN. USE ENTER KEY TO CHANGE THE 

SETTING 
 
 

Summary of Contents for PCE-TG 250

Page 1: ...1 0 mm TO 250 mm IN STEEL DEPENDING OF THE PROBE THROUGH COATING 4 mm 18 mm TOLERANCE 0 02 mm IN STEEL DISPLAY RESOLUTION 0 01 mm Less than 100 0 mm 0 1 mm over 100 mm BULIT IN STANDARD BLOCK 4 00 mm...

Page 2: ...PLANT INDICATOR AUTOMATIC POWER OFF DEVICE HIGH LOW LIMIT ALARM LCD BRIGHTNESS ADJUST THROUGH COATING CAPABILTY THROUGH COATING MEASUREMENT 500 DATA READING STORAGE WITH DIFFERENT 5 MEMORY FILES LOW B...

Page 3: ...WILL DISPLAY PRESS MENU KEY 2 TIME THEN PRESS ENTER KEY DISPLAY FOLLOWING System Setup Function Setup Outside alarm enter to select to down PRESSS ENTER KEY AGAIN DISPLAY FOLLOWING SCREEN Units METRIC...

Page 4: ...HT CURSOR TO FUNCTION SETTING PRESS THEN USE KEY MOVE HIGH LIGHT CURSOR DOWN TO THE FOLLOWING DISPLAY Erase file Erase all data Erase cal data ENTER to select UP AND DOWN LIMIT ALARM SETTING PRESS GAU...

Page 5: ...Y IS NOT REQUIRED THEN PRESS VEL KEY SELECT THE VELOCITY FROM MEMORY THEN PRESS ENTER KEY IF THE 5 STORGED VEL ARE NOT THE TEST MATERIAL VEL THEN USE KEY ADJUST THE VEL DISPLAY UNTIL MEET REQUIRED VEL...

Page 6: ...THE GAUGE SIDE BLOCK PRESS THE PROBE TO THE BLOCK WAIT UNTIL COUPLANT INDICATOR NOT FLASH PRESS CAL KEY THE DISPLAY WILL SHOW 4 00 mm TWO POINT CALIBRATION TO GET MORE PRECISION MENSURMENT TWO POINT...

Page 7: ...F INTO ON Resolution HIGH Min Captyre OFF 2 Point CAL ON enter to select TO UP to down PRESS MANUAL BACK THEN PRESS KEY AGAIN TO THE DISPLAY SCREEN PUT COUNPLAT ON THIN BLOCK MOVE THE PROBE ON THE THI...

Page 8: ...THE PROBE THE THICKNESS DISPLAY NO CHANGE BUT COUPLANT INDICTOR DISAPPEAR 5 LOW BATTERY INDICATOR WHEN THICKNESS GAUGE POWER WAS NOT ENOUGH THE LOW BATTERY INDICATOR WILL DISPLAY ON THE LEFT BOTTOM O...

Page 9: ...DE WILL SHOW MEMORY THIS MEAN THE READING 1 ALREADY STORAGED IN THE FILE 1 CONTINOUS TO DO THE STORAGE OF READING 2 3 4 7 2 THCINESS DATA RECALL PRESS KEY MOVE DISPLAY LCD HIGH LIGHT CUSOR TO SAVE1 PR...

Page 10: ...ILE PRESS GAUGE KEY TURN ON THE GAUGE LCD WILL DISPLAY PRESS ENTER 2 TIMES PRESS ENTRY KEY USE KEY MOVE HIGH LIGHT CURSOR TO FUNCTION SETTING PRESS THEN USE KEY MOVE HIGH LIGHT CURSOR DOWN TO THE FOLL...

Page 11: ...2D CAN MEASURMENT THE THICKNESS WITHOUT MOVE THE COATING PRESS KEY TURN ON THE GAUGE PRESS KEY MOVE THE HIGH LIGHT CURSOR TO Std PRESS KEY TO CHANGE INTO Coat NOW THE GAUGE UNDER THOUGH COATING MEASUR...