GRL-PCIE5-CEM-RXA User Guide and MOI
Rev7.0
© Granite River Labs 2022 Version 7.0, June 2022. Updated 06.29.2022
Page 63 of 123
8.1.7
Preset Calibration Adjustment
Adjust the Pre and Post Cursor1 of Presets 0 to 9 with Scope markers using the following
configuration.
Preset 4 Calibration:
a)
BERT Settings:
•
General Output: ON
•
Select SI-PPG and Emphasis tab- Emphasis Function: ON
•
Select the Pattern tab- Test Pattern: Using Preset 4, with 64 ones followed by 64 zeros
followed by 128 bits of a 1010 clock pattern at 32.0 GT/s
•
Select the Misc1 tab- Aux Output: Pattern Sync