GRL-PCIE5-CEM-RXA User Guide and MOI
Rev7.0
© Granite River Labs 2022 Version 7.0, June 2022. Updated 06.29.2022
Page 85 of 123
4.
In the MX183000A - Selector pop-
up, select ‘PCIe Link Training’ and then ‘Start’.
5.
In the next window, select the network address of the MX190000A BERT application followed
by ‘Connect’ to link up with the MX190000A.
6.
In the MX183000A - PCIe Link Training window, select the Link Training tab. Select the
“Setting” check box on the right
of the screen.
7.
When the LEQ Test pane appears, select the Initial Tx LEQ tab. Set the
“DUT Initial Preset
(Preset Hint Tx)” to Preset 0
(P0).
Then select “Apply”.
8.
Press the Power Reset button on the CBB test fixture to reset the DUT.