background image

Simple user maintenance guide

Desiccant capsule replacement 

When to replace?

 Silica gel will darken as it becomes saturated. 

Depending on usage a 12 week minimum lifetime is expected.

Step 1:

 Check Silica gel colour 

via the window on the 
left-hand side panel of the 
generator (if blue/green it 
needs to be replaced).

Step 2: 

Press and hold the 

power button for 5 seconds to 
shutdown and depressurize. 
Now turn off the main power 
switch at the back of the unit. 

Step 4:

 Loosen and remove the 

cap of the desiccant dryer by 
placing the tool onto the cap 
and turning counter-clockwise.

Step 5:

 Remove the saturated 

desiccant capsule and replace 
with a fresh capsule.

Step 6:

 Replace the cap and place 

the tool on top, tighten with the 
tool by turning it clockwise.

Step 8: 

Power up the generator using the main power switch 

on the rear of the unit. Once diagnostics are complete, the 
LEDs will be solid green and the generator will be ready to 
deliver gas to the application. Press the front button to 
supply gas to the application.

Precision SL A5 Quick INS  Rev 

2

 PN 3303233  RSID 3374

Step 7: 

Close the top panel.

Step 3:

 Open the top panel by 

lifting gently.

Reviews: