background image

PEAK 

Scientific Instruments Ltd 

High Purity Nitrogen Generator 

Instructions for use Manual 

20

th

 September 2007 

 

NM20ZL User Guide Issue 2

 

Commissioning

 

 

With the Generator installed as described earlier connect compressed air to the unit and turn it ON. 

 Select Source N2 or Source Air mode via selector switch. Disconnect the Outlet connections to allow the 

generator to vent to atmosphere until the unit is stabilised. The Generator has been pre-set in the factory 

to give the specified output flow-rate and pressure in line with known requirements. Once the Membranes 

reach  the  design  pressures  the  Generator  will  stabilise  and  produce Nitrogen. Maximum purity will be 

achieved. The generator can then be re-connected to the application. 

 

The design of the generator is such that it can deliver up to 26 Litres / min Nitrogen/ Air at 110 psig with 

an additional 12 Litres / min of Nitrogen at 80 psig. Should the demand for gases be less than the rated 

output  flow,  or  indeed  should  the  demand  stop  the  generator  will  continue  to  operate  without  any 

problems.

 

 

Pressure & Flow Settings 

 

Output pressures and flow rates are controlled by pressure regulators located behind the front panel as 

shown below. The Factory settings for the 3 outputs are as follows: - 

 

 

 

 

 

  

Port 

Usage / Connection 

Pressure / Flow 

CURTAIN 

Curtain Gas 

 12 ltr /min - 80 psig 

SOURCE 

Source Air / Source N2  

  26 ltr/min - 110 psig 

EXHAUST 

Exhaust Gas 

  8 ltr/min - 60 psig 

 

The  above  settings  should  allow  the  NM20ZL  to  be  operated  with  all  standard  configurations  for  the 

Applied Biosystems Api range of LC/MS/MS instruments. Should the above settings not provide sufficient 

flow or pressure for your application please contact the factory for assistance. 

Pressure Regulators 

Flow Controllers 

Summary of Contents for NM20ZL

Page 1: ...ain Crescent Inchinnan Business Park Inchinnan Renfrew PA4 9RE Scotland UK Tel 44 0 141 812 8100 Fax 44 0 141 812 8200 info peakscientific com www peakscientific com USA 19 Sterling Road Suite 8 Bille...

Page 2: ...gen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 2 Document Change History Rev No Changed Initials Date 1 Document Created SK 20 09 07 2 Included new design of c...

Page 3: ...Allow the company to inspect the Goods and or any installation and any relevant packaging as and when reasonably required by the Company 5 Subject to the Company being notified of any defect as is ref...

Page 4: ...ZL User Guide Issue 2 4 Contents Page No Document History 2 Warranty Statement 3 Safety Notice To Users 3 Introduction 5 Unpacking Installation 5 Location 5 Air Connection 6 Principle of Operation 7 C...

Page 5: ...ectrometer it is supplying If this is not convenient then the generator can be sited elsewhere however consideration should be made to the lengths of pipe runs as pressure drops can result from extend...

Page 6: ...the unit Fittings are provided to connect these ports to the gas connections of the Mass Spectrometer A kit of fittings for connecting to the Mass Spectrometer is provided with this User Guide LC MS w...

Page 7: ...oves bulk moisture and harmful particles The air is now divided into 2 streams The first stream is passed through the Nitrogen Membrane to generate a flow of Nitrogen for the Curtain Gas output The Ni...

Page 8: ...or is such that it can deliver up to 26 Litres min Nitrogen Air at 110 psig with an additional 12 Litres min of Nitrogen at 80 psig Should the demand for gases be less than the rated output flow or in...

Page 9: ...ver as with all scientific and technical equipment it should be regularly inspected and serviced as below BA Inlet Filters Every 12 months RAC Filter Elements Every 12 months 1 BA inlet Filter Element...

Page 10: ...ter The Element is as shown and should be changed at 12 month intervals The Air supply to Generator MUST be turned OFF and the Generator MUST be de pressurised prior to attempting to remove ANY filter...

Page 11: ...psig 10 0 Barg Minimum Air Inlet Flow Rate 90 l min ATP Outlet Gas Maximum Pressure Drop Outlet Inlet P 8 psig Maximum Gas Outlet Pressure 110 psig Maximum Source Gas Outlet Flow Nitrogen Air 26 Litre...

Page 12: ...PEAK Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 12 9 Pneumatic Diagram...

Page 13: ...Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 13 Maintenance Log Model NM20ZL Serial number Work Done Remarks Date...

Page 14: ...PEAK Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 14 Notes...

Reviews: