WARNING:
Servicing and/or repair of the Generator should only be undertaken by a
TECHNICALLY COMPETENT PERSON with the Generator in a safely isolated condition.
Due to the simplicity of the design and the small number of moving parts the NM Series Nitrogen
Generator will have a long and trouble free life. However as with all scientific and technical equipment it
should be regularly inspected and serviced as below
BA Inlet Filters
Every 12 – months
RAC Filter Elements Every 12 - months
BA Inlet Filter Elements
These should be changed at 12 months intervals. In addition filter bowls should be cleaned and, the
operation of the auto-drains should be checked, if auto-drain requires replacing. Part # is (02-4485).
FAILURE TO FOLLOW THE PRESCRIBED MAINTENANCE PLAN WILL
INVALIDATE THE PRODUCT WARRANTY
The Air supply to the Generator MUST be turned OFF and the Generator MUST
be de-pressurised prior to attempting to remove ANY filter bowl. Failure to do
this may cause injury.