50
* When you short the test leads in the 200
range, your
meter displays a small value (no more than 0.3
). This
value is due to your meter’s and test leads internal
resistance. Make a note of this value and subtract it from
small resistance measurements for better accuracy.
* If in 200
, please short the test leads and measure wine
resistance. And then subtract the resistance from the
value measured.
* It is normal to display 10M
when the test leads shorted
in range 2000M
, it will not affect the accuracy and shall
be subtracted from the value measured. For example:
The object resistance is 1000 M
, the reading value is
1010M
, then the correct value shall be
1010M
– 10M
= 1000M
.