2. Preparing of operation
2.1. Using the test leads
Use only the type of test leads supplied with your meter. These test leads are rated for 1000 Volts.
Warning!
* If you try to measure DC voltages above 1000 V or AC voltages above 750 V AC, you might damage your
meter and expose yourself to a serious shock hazard. Use extreme care when you measure high voltages.
* Never connect the test lead plug into the COM terminal to a source of voltage greater than 500 Volts with
respect to earth ground. This creates a serious shock hazard.
3. Specifications
3.1. General Characteristics
display
21999 counts (4 ½-digit) with automatic polarity indication
Reading times
3 - 4 x per second
Operation temperature 0 … 40°C
Temperature for
guaranteed Accuracy +23°C ± 5°C
Storage Temperature -10°C … + 50°C
Operation voltage 9 V battery (NEDA 1604, 6F22))
Dimensions (WxHxD) 210 x 92 x 40 mm
Weight
410 g (including battery)
3.2. Ranges
DC Voltage
Range
Resolution
Accuracy
220 mV
100 µV
± 0,12% of rdg. + 3
dgt.
± 0,12% of rdg. + 5
dgt.
2,2 V
1 mV
22 V
10 mV
220 V
100 mV
1000 V
1 V
± 0,5% of rdg. + 5
dgt.
AC Voltage
Range
Resolution
Accuracy
220 mV
100 µV
± 1,0% of rdg. + 3
dgt.
(50 Hz ~ 5 kHz)
± 2,5% of rdg. + 10
dgt.
(5 kHz ~ 20 kHz)
2,2 V
1 mV
22 V
10 mV
220 V
100 mV
750 V
1 V
± 0,5% of rdg. + 10
dgt.
(50 Hz ~ 400 Hz)