3.9. Temperature
Range
Resolution
Accuracy
-55°C~ 0°C
0,1°C
5,0% rdg.+ 4°C
1°C~ 400°C
2,0% rdg.+ 3°C
401°C~1000°C
1,0°C
2,0% rdg.
Overload protection: 250 V DC/220 V AC
rms
3.10. Continuity Test
Buzzer beeps continuously at less than 30
.
Overload protection: 250 V DC/AC
rms
Open circuit voltage: 0,5 V